CLIC ABSORBER 800+ MX00004

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2012-01-24 for CLIC ABSORBER 800+ MX00004 manufactured by Draeger Medical, Inc..

Event Text Entries

[2511118] A fresh co2 absorber canister was installed in the anesthesia machine. Anesthesia machine checked ok, but during use it continually announced a "fresh gas low" alarm. Clinician was able to adjust machine settings to maintain correct patient ventilation. When case was ended, problem was reported to biomed techs. Upon investigation, a fault within the absorber canister was identified which allowed soda lime pellets to partially fill the canister's central air inlet tube, greatly reducing air flow. The anesthesia machine worked properly once a new canister was installed.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2430881
MDR Report Key2430881
Date Received2012-01-24
Date of Report2012-01-24
Date of Event2012-01-20
Report Date2012-01-24
Date Reported to FDA2012-01-24
Date Added to Maude2012-01-30
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameCLIC ABSORBER 800+
Generic NameCO2 ABSORBER CARTRIDGE, 1.2L DISPOSABLE
Product CodeBSF
Date Received2012-01-24
Model Number800+
Catalog NumberMX00004
Lot NumberARCM F006
ID Number*
OperatorPHYSICIAN
Device Sequence No1
Device Event Key0
ManufacturerDRAEGER MEDICAL, INC.
Manufacturer Address3135 QUARRY RD TELFORD PA 18969 US 18969


Patients

Patient NumberTreatmentOutcomeDate
10 2012-01-24