IMMAGE IMMUNOCHEMISTRY SYSTEM RHEUMATOID (RF) REAGANT 447070

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,05,06 report with the FDA on 2012-01-30 for IMMAGE IMMUNOCHEMISTRY SYSTEM RHEUMATOID (RF) REAGANT 447070 manufactured by Beckman Coulter, Inc..

Event Text Entries

[18178502] The customer reported false positive rheumatoid factor (rf) patient results with the use of a specific immage immunochemistry system rheumatoid factor (rf) reagent lot on an immage immunochemistry system. The customer reran the samples using a different rf reagent lot and obtained lower, negative rf results. The number of patients involved in this event is unknown. The customer did not provide actual patient results. The erroneous results were not reported out of the laboratory and hence there were no reports of death, serious injury or change to patient treatment associated or attributed to this event. There were no reports of any issues with other chemistries or system errors. The samples were fresh, serum samples which were centrifuged prior to testing. No sample issues were noted. Instrument chemistry quality control results shifted high during the timeframe of the event. No patient information or additional system performance information was provided by the customer.
Patient Sequence No: 1, Text Type: D, B5


[18521820] Service was not dispatched for this incident as the issue is believed to be reagent related.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2050012-2012-00240
MDR Report Key2432360
Report Source01,05,06
Date Received2012-01-30
Date of Report2012-01-04
Date of Event2012-01-03
Date Mfgr Received2012-01-04
Device Manufacturer Date2010-12-16
Date Added to Maude2012-08-09
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS DUNG NGUYEN
Manufacturer Street250 S KRAEMER BLVD
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal92821
Manufacturer Phone7149614941
Manufacturer G1BECKMAN COULTER, INC.
Manufacturer Street250 S KRAEMER BLVD.
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal Code92821
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameIMMAGE IMMUNOCHEMISTRY SYSTEM RHEUMATOID (RF) REAGANT
Generic NameSYSTEM, TEST, RHEUMATOID FACTOR
Product CodeDHR
Date Received2012-01-30
Model NumberNA
Catalog Number447070
Lot NumberM008778
ID NumberNA
Device Expiration Date2012-12-31
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerBECKMAN COULTER, INC.
Manufacturer Address250 S KRAEMER BLVD. BREA CA 92821 US 92821


Patients

Patient NumberTreatmentOutcomeDate
10 2012-01-30

© 2024 FDA.report