IMMAGE? 800 IMMUNOCHEMISTRY SYSTEM IMMAGE 800 A15445

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2012-01-31 for IMMAGE? 800 IMMUNOCHEMISTRY SYSTEM IMMAGE 800 A15445 manufactured by Beckman Coulter Inc..

Event Text Entries

[2517277] Customer called on (b)(6) 2012 reporting of a leak coming from the sample probe of an immage 800 immunochemistry system. Customer stated that the instrument was serviced the day before the incident and all qc and patient samples were good when the field service engineer (fse) had left the site. Customer stated that no patient samples were run and no erroneous results were generated as a result of the leak. Customer was wearing personal protective equipment at the time of the leak and no exposure or injury was reported. Fse returned on 01/13/2012 and replaced the sample syringe assembly with a new assembly. Issue was resolved and repair was verified as per established procedure.
Patient Sequence No: 1, Text Type: D, B5


[9589103] Evaluation - results: fse returned the sample syringe assembly. (b)(4)..
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2050012-2012-00312
MDR Report Key2433877
Report Source05,06
Date Received2012-01-31
Date of Report2012-01-12
Date of Event2012-01-12
Date Mfgr Received2012-01-12
Device Manufacturer Date2004-07-20
Date Added to Maude2012-08-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS. DUNG NGUYEN
Manufacturer Street250 S. KRAEMER BOULEVARD
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal92821
Manufacturer Phone7149614941
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameIMMAGE? 800 IMMUNOCHEMISTRY SYSTEM
Generic NameNEPHELOMETER, FOR CLINICAL USE
Product CodeJQX
Date Received2012-01-31
Model NumberIMMAGE 800
Catalog NumberA15445
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBECKMAN COULTER INC.
Manufacturer Address250 S. KRAEMER BLVD BREA CA 92821 US 92821


Patients

Patient NumberTreatmentOutcomeDate
10 2012-01-31

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.