IMMAGE? 800 IMMUNOCHEMISTRY SYSTEM IMMAGE 800 A15445

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2012-02-01 for IMMAGE? 800 IMMUNOCHEMISTRY SYSTEM IMMAGE 800 A15445 manufactured by Beckman Coulter, Inc..

Event Text Entries

[2555814] Customer reported erroneous low test results for igg (immunoglobulin g), iga (immunoglobulin a), igm (immunoglobulin m), kappa and lambda were obtained for one patient sample when using the immage 800 immunochemistry system. The erroneous test results were not reported out of the laboratory. Quality control data provided by the customer demonstrated imprecision. There were no reports of death, serious injury or affect to patient treatment associated with this event.
Patient Sequence No: 1, Text Type: D, B5


[9623078] A beckman coulter field service engineer (fse) evaluated the analyzer and replaced the sample probe. The fse performed sample crane alignments, verified both syringes and cuvettes, and ran multiple patient samples and quality control. The probable cause of this event was attributed to a malfunction of the sample probe. This mdr is one of two events reported by this customer. The related mdr is 2050012-2012-00338.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2050012-2012-00337
MDR Report Key2433960
Report Source05,06
Date Received2012-02-01
Date of Report2012-01-11
Date of Event2012-01-09
Date Mfgr Received2012-01-11
Device Manufacturer Date2007-12-19
Date Added to Maude2012-08-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS DUNG NGUYEN
Manufacturer Street250 S KRAEMER BLVD
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal92821
Manufacturer Phone7149614941
Manufacturer G1BECKMAN COULTER, INC.
Manufacturer Street250 S KRAEMER BLVD
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal Code92821
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameIMMAGE? 800 IMMUNOCHEMISTRY SYSTEM
Generic NameNEPHELOMETER, FOR CLINICAL USE
Product CodeCFF
Date Received2012-02-01
Model NumberIMMAGE 800
Catalog NumberA15445
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBECKMAN COULTER, INC.
Manufacturer Address250 S KRAEMER BLVD BREA CA 92821 US 92821


Patients

Patient NumberTreatmentOutcomeDate
10 2012-02-01

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.