IMMULITE 2000 INSULIN L2KIN2

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2012-02-01 for IMMULITE 2000 INSULIN L2KIN2 manufactured by Siemens Healthcare Diagnostics.

Event Text Entries

[2468536] Discordant low insulin results were obtained for patient samples on an immulite 2000 analyzer. The customer performed a (b)(4) study with their new lot of insulin reagent (lot 314) and their prior lot of insuliln reagent (lot 308). The customer reported increased imprecision and a negative bias with insulin reagent lot 314 (compared to lot 308). There were no known reports of patient treatment being altered or prescribed due to the discordant low insulin results. There were no known reports of adverse health consequences due to the discordant low insulin results.
Patient Sequence No: 1, Text Type: D, B5


[2603754] Discordant low insulin results were obtained for patient samples on an immulite 2000 analyzer. The customer performed a correlation study with their new lot of insulin reagent (lot 314) and their prior lot of insulin reagent (lot 308). The customer reported increased imprecision and a negative bias with insulin reagent lot 314 (compared to lot 308). There were no known reports of patient treatment being altered or prescribed due to the discordant low insulin results. There were no known reports of adverse health consequences due to the discordant low insulin results.
Patient Sequence No: 1, Text Type: D, B5


[9593132] The customer contacted the siemens technical solutions center (tsc) about the discordant low insulin results with lot 314. The cause of the discordant low insulin results is unknown at this time. This incident is currently under investigation.
Patient Sequence No: 1, Text Type: N, H10


[9837312] Siemens filed the initial mdr on 02/01/2012. Siemens healthcare dx has implemented a field correction (urgent device recall ur005-2012-04-26, april 2012) for the immulite 2000/2000 xpi insulin assay lots 312-316.
Patient Sequence No: 1, Text Type: N, H10


[9959080] The customer contacted the siemens technical solutions center (tsc) about the discordant low insulin results with lot 314. The cause of the discordant low insulin results is unknown at this time. This incident is currently under investigation. (b)(4) 2012 additional information: the immulite 2000 insulin imprecision on lot# 314 has been investigated by siemens. Siemens has determined that the cause of the immulite 2000 insulin imprecision on lot# 314 is due to a reduction in the concentration of the insulin bead coat antibody (raw material). The device is running within specifications. No further evaluation of the device is necessary.
Patient Sequence No: 1, Text Type: N, H10


[10060003] The 510(k) number has been corrected: k022603.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2432235-2012-00028
MDR Report Key2435600
Report Source05,06
Date Received2012-02-01
Date of Report2012-01-05
Date of Event2012-01-05
Date Mfgr Received2012-06-07
Date Added to Maude2012-07-31
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactLORIANN RUSSO
Manufacturer Street511 BENEDICT AVENUE
Manufacturer CityTARRYTOWN NY 10591
Manufacturer CountryUS
Manufacturer Postal10591
Manufacturer Phone9145242287
Manufacturer G1SIEMENS HEALTHCARE DIAGNOSTICS
Manufacturer StreetGLYN RHONWY LLANBERIS
Manufacturer CityCAERNATION, GWYNEDD, WALES LL554EL
Manufacturer CountryUK
Manufacturer Postal CodeLL55 4EL
Single Use3
Previous Use Code3
Removal Correction NumberUR005-2012-04-26
Event Type3
Type of Report3

Device Details

Brand NameIMMULITE 2000 INSULIN
Generic NameINSULIN ASSAY
Product CodeCFP
Date Received2012-02-01
Model NumberIMMULITE 2000 INSULIN
Catalog NumberL2KIN2
Lot Number314
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS
Manufacturer AddressGLYN RHONWY LLANBERIS GWYNEDD, WALES LL554EL UK LL55 4EL


Patients

Patient NumberTreatmentOutcomeDate
10 2012-02-01

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.