NASAL FORCEPS 8211.101

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 00,06 report with the FDA on 2012-01-20 for NASAL FORCEPS 8211.101 manufactured by Richard Wolf Gmbh.

Event Text Entries

[2517808] The pituitary forceps used during a procedure broke apart while in use. What was the original intended procedure: bilateral orbital decompression. Facility e-mail date (b)(6) 2012: "no harm to patient. " facility also noted that the technician noticed a pin missing when the surgeon handed back the instrument, the eye was checked and the missing pin was not found. Facility e-mail dated (b)(6) 2012: "pin was missing from the tip of the instrument, near the jaw. "
Patient Sequence No: 1, Text Type: D, B5


[9616438] Facility did not return the device to rwmic for analysis, instead it was shipped to a 3rd party service provider for repair. Rwmic records were reviewed, this facility has not used rwmic as a service provider for this device since distribution. Instrument age: 8+ years. Distribution date: august 21, 2003. Possible root cause: improper maintenance (none, previous 3rd party repair) and/or pin came loose during reprocessing or normal use. Reprocessing as reported by facility: "all instruments that can be processed by machine first go thru a 10 min ultrasonic cleaning with a neutral ph cleaning agent. " after this process it goes thru the washer at 170 degrees for up to 45 minutes' it is then visually inspected, placed in the acceptable container or wrap paper and sterilized at a 10 min / 40 min sterilization / dry time at 274 degrees temp. " labeling for this device was reviewed, instruction g56 includes the following: important notifications: "to ensure maximum safety, we strongly recommend that this equipment be sent to our facility periodically, depending on its use, for comprehensive inspection. " "richard wolf, as manufacturer and seller of the instruments which are the subject of this manual, is not responsible for any direct or indirect damages resulting from improper or incorrect use, care, or servicing of these instruments, or from the failure to follow the instructions in this manual. " proper care and maintenance is important for the efficient and safe operation of sophisticated medical / surgical equipment. We recommend careful inspection of all equipment upon receipt and prior to each use as a safeguard against possible injury to patient and/or operator. When in doubt as to proper care and maintenance procedures, contact the richard wolf customer service department at (b)(4) or contact your local richard wolf representative. " "if non-authorized persons perform repairs or modifications of this product, richard wolf does not accept liability and the warranty becomes void. Components influencing safety may only be replaced with original replacement parts. " care and maintenance: "ultrasonic cleaning is recommended for all hand instruments with the following cautions: the vibration associated with ultrasonic cleaning may loosen lock screws on instruments that have them. " rwmic considers this matter closed, but will provide additional / follow-up information to fda if received or requested. Rwmic will report our finding to facility, we will also include a copy of instruction g56. Facility will be cautioned not to use 3rd party repair for "pin replacement," etc.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1418479-2011-00026
MDR Report Key2436597
Report Source00,06
Date Received2012-01-20
Date of Report2012-01-19
Date of Event2011-12-07
Date Facility Aware2011-12-22
Report Date2012-01-18
Date Reported to FDA2012-01-19
Date Reported to Mfgr2012-01-18
Date Mfgr Received2012-01-18
Device Manufacturer Date2002-09-01
Date Added to Maude2012-02-03
Event Key0
Report Source CodeDistributor report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationRISK MANAGER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactRON HASELHORST
Manufacturer Street353 CORPORATE WOODS PARKWAY
Manufacturer CityVERNON HILLS IL 60061
Manufacturer CountryUS
Manufacturer Postal60061
Manufacturer G1RICHARD WOLF GMBH
Manufacturer CityKNITTLINGEN
Manufacturer CountryGM
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameNASAL FORCEPS
Generic NameNASAL FORCEPS
Product CodeKBE
Date Received2012-01-20
Model Number8211.101
Catalog Number8211.101
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Age8 YR
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerRICHARD WOLF GMBH
Manufacturer AddressKNITTLINGEN GM


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2012-01-20