[2556860] Customer reported that while cutting with the microtome, the hand grip of the hand wheel of the rotary microtome came loose from the bolt on which the hand grip is mounted.
Patient Sequence No: 1, Text Type: D, B5


[9593691] The investigation indicated the cause for the malfunction was due to the incorrect functioning of the handwheel handle. The hand grip which were mounted on a bolt were not properly held, because of not enough glue between the bold and hand grip was applied. The handwheel was completer replaced, anda performance test was conducted to ensure that the device was working properly.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number8010478-2011-00012
MDR Report Key2437477
Report Source05
Date Received2011-12-27
Date of Report2011-11-23
Date of Event2011-11-16
Date Mfgr Received2011-11-23
Device Manufacturer Date2011-10-01
Date Added to Maude2012-07-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactSTEFFAN LAABS
Manufacturer StreetHEIDELBERGER STR. 17-19
Manufacturer CityNUSSLOCH 69226
Manufacturer CountryGM
Manufacturer Postal69226
Manufacturer Phone224143345
Single Use3
Remedial ActionRP
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameLEICA RM2235
Generic NameMICROTOME
Product CodeIDO
Date Received2011-12-27
Model NumberRM2235
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerLEICA BIOSYSTEMS GMBH
Manufacturer AddressHEIDELBERGER STR. 17-19 NUSSLOCH 69226 GM 69226


Patients

Patient NumberTreatmentOutcomeDate
10 2011-12-27