IMMAGE ALPHA-1-ANTITRYPSIN (AAT) 447740

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 02,05,06 report with the FDA on 2012-02-07 for IMMAGE ALPHA-1-ANTITRYPSIN (AAT) 447740 manufactured by Beckman Coulter, Inc..

Event Text Entries

[9957254] Evaluation- method/conclusions: device from the same lot was evaluated. The evaluation confirmed the imprecision issue at lower concentrations of aat.
Patient Sequence No: 1, Text Type: N, H10


[19929522] Customer reported imprecision with the immage alpha-1-antitrypsin (aat) reagent when using the immage 800 immunochemistry system. The imprecision was identified with the quality control samples with lower concentrations of aat. While there were no reports of erroneous patient test results, a similar difference in test results could occur with patient samples. There were no reports of death or serious injury associated with this event. It is unknown if there was any affect to patient treatment associated with this event.
Patient Sequence No: 1, Text Type: D, B5


[20278390] This issue is under investigation.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2050012-2012-00339
MDR Report Key2440998
Report Source02,05,06
Date Received2012-02-07
Date of Report2012-01-12
Date of Event2012-01-12
Date Mfgr Received2012-04-03
Device Manufacturer Date2011-04-22
Date Added to Maude2012-08-14
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS. DUNG NGUYEN
Manufacturer Street250 S KRAEMER BLVD.
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal92821
Manufacturer Phone7149614941
Manufacturer G1BECKMAN COULTER, INC.
Manufacturer Street250 S KRAEMER BLVD
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal Code92821
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameIMMAGE ALPHA-1-ANTITRYPSIN (AAT)
Generic NameALPHA-1-ANTITRYPSIN, ANTIGEN, ANTISERUM, CONTROL
Product CodeDEM
Date Received2012-02-07
Model NumberNA
Catalog Number447740
Lot NumberM105003
ID NumberNA
Device Expiration Date2013-03-31
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerBECKMAN COULTER, INC.
Manufacturer Address250 S KRAEMER BLVD BREA CA 92821 US 92821


Patients

Patient NumberTreatmentOutcomeDate
10 2012-02-07