SIMVIEW 3000 SIMULATOR 01957153 NA

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 07 report with the FDA on 2012-01-19 for SIMVIEW 3000 SIMULATOR 01957153 NA manufactured by Siemens Medical Solutions Usa, Inc., Radiation Oncology.

Event Text Entries

[20739581] Siemens received a report on (b)(6) 2011, noting that when the operator moved the table down at the end of a patient's simulation session on the simview 3000 simulator, the table rapidly fell down about 50 cm with the patient on top of it. Reportedly, the patient was not injured. The patient came down from the table and went out of the room on his own. Since that time the table did not move in the vertical direction. This reported issue occurred in (b)(6).
Patient Sequence No: 1, Text Type: D, B5


[20907673] Siemens became aware of the reported event on (b)(4) 2011 and are mailing the mdr on (b)(4) 2012. Siemens customer service engineer found during evaluation that the main chain for the motor and the lifting screw or barring were both broken. The broken chain was replaced, and a new table base and table cable were installed. The simulator is back in use by the customer. (b)(6).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2910081-2012-00006
MDR Report Key2443390
Report Source07
Date Received2012-01-19
Date of Report2011-12-20
Date of Event2011-12-20
Date Mfgr Received2011-12-20
Date Added to Maude2012-08-08
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMARLYNNE GALLOWAY
Manufacturer Street51 VALLEY STREAM PARKWAY D02
Manufacturer CityMALVERN PA 19355
Manufacturer CountryUS
Manufacturer Postal19355
Manufacturer Phone6102195361
Manufacturer G1SIEMENS MEDICAL SOLUTIONS USA, INC.
Manufacturer Street4040 NELSON AVE.
Manufacturer CityCONCORD CA 94520
Manufacturer CountryUS
Manufacturer Postal Code94520
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSIMVIEW 3000 SIMULATOR
Generic NameSYSTEM, SIMULATION, RADIATION THERAPY
Product CodeKPQ
Date Received2012-01-19
Model Number01957153
Catalog NumberNA
Lot NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS MEDICAL SOLUTIONS USA, INC., RADIATION ONCOLOGY
Manufacturer Address4040 NELSON AVE. CONCORD CA 94520 US 94520


Patients

Patient NumberTreatmentOutcomeDate
10 2012-01-19

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.