HYPERGLIDE OCCLUSION BALLOON SYSTEM 104-4112

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 04,05,06 report with the FDA on 2012-02-09 for HYPERGLIDE OCCLUSION BALLOON SYSTEM 104-4112 manufactured by Ev3 Neurovascular.

Event Text Entries

[20794215] It was reported the balloon could not be visualized during contrast injection (50:50 ratio). The physician decided to used 90% concentration contrast media to inflate the balloon and it could not be deflated. The inflated balloon was successfully removed from the patient. No patient injury reported.
Patient Sequence No: 1, Text Type: D, B5


[20965897] The balloon catheter has been evaluated and a large amount of blood was found in the balloon lumen. Based upon the findings, the large amount of blood found in the balloon assembly was the likely cause of the non-deflation. The instruction for use warns "inadvertent proximal guidewire movement can cause blood to enter the balloon lumen which may inhibit balloon deflation. "(b)(4).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2029214-2012-00046
MDR Report Key2445008
Report Source04,05,06
Date Received2012-02-09
Date of Report2012-01-11
Date of Event2012-01-11
Date Mfgr Received2012-01-11
Device Manufacturer Date2010-01-18
Date Added to Maude2012-08-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactHENRY TO
Manufacturer Street9775 TOLEDO WAY
Manufacturer CityIRVINE CA 92618
Manufacturer CountryUS
Manufacturer Postal92618
Manufacturer Phone9496801335
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameHYPERGLIDE OCCLUSION BALLOON SYSTEM
Generic NameOCCLUSION BALLOON CATHETER
Product CodeNUF
Date Received2012-02-09
Returned To Mfg2012-01-26
Model Number104-4112
Lot Number8119140
Device Expiration Date2012-11-01
OperatorPHYSICIAN
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerEV3 NEUROVASCULAR
Manufacturer Address9775 TOLEDO WAY IRVINE CA 92618 US 92618


Patients

Patient NumberTreatmentOutcomeDate
10 2012-02-09

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.