IMMAGE? RHEUMATOID FACTOR REAGENT IMMAGE RF 447070

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,05,06 report with the FDA on 2012-02-09 for IMMAGE? RHEUMATOID FACTOR REAGENT IMMAGE RF 447070 manufactured by Beckman Coulter, Inc..

Event Text Entries

[15507483] The customer reported false positive rheumatoid factor (rf) results were generated on the immage rheumatoid factor reagent lot m012376 which used on immage immunochemistry system. The customer stated that the number of rf high results have increased since the use of rf reagent lot m012376. The customer also stated that qc results have increased. The results were not reported out of the laboratory; hence patient treatment was not impacted by this event. The customer recalibrated the instrument, however, the issue remained. The instrument is currently performing within qc specifications with respect to controls (accuracy) and reproducibility (precision).
Patient Sequence No: 1, Text Type: D, B5


[15693654] Root cause is not known but this appears to be a reagent related issue. The customer was sent a replacement lot of reagent.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2050012-2012-00429
MDR Report Key2445150
Report Source01,05,06
Date Received2012-02-09
Date of Report2012-01-19
Date of Event2012-01-17
Date Mfgr Received2012-01-19
Device Manufacturer Date2011-03-31
Date Added to Maude2012-08-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMRS. DUNG NGUYEN
Manufacturer Street250 S. KRAEMER BLVD.
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal92821
Manufacturer Phone7149614941
Manufacturer G1BECKMAN COULTER, INC.
Manufacturer Street2470 FARADAY AVE.
Manufacturer CityCARLSBAD CA 92010
Manufacturer CountryUS
Manufacturer Postal Code92010
Single Use3
Previous Use Code3
Removal Correction Number2050012-01/26/2012-003C
Event Type3
Type of Report3

Device Details

Brand NameIMMAGE? RHEUMATOID FACTOR REAGENT
Generic NameSYSTEM, TEST, RHEUMATOID FACTOR
Product CodeDHR
Date Received2012-02-09
Model NumberIMMAGE RF
Catalog Number447070
Lot NumberM012376
ID NumberN/A
Device Expiration Date2013-03-31
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerBECKMAN COULTER, INC.
Manufacturer Address250 S KRAEMER BLVD BREA, CA 92821 US 92821


Patients

Patient NumberTreatmentOutcomeDate
10 2012-02-09

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.