BOND AND NOVOCASTRA PA0271,PA0266, CD7 PA0271, PA0266, CD7

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 07 report with the FDA on 2012-02-06 for BOND AND NOVOCASTRA PA0271,PA0266, CD7 PA0271, PA0266, CD7 manufactured by Leica Biosystems Newcastle Ltd.

Event Text Entries

[2537587] (b)(4) was received from a customer in (b)(6) regarding bond rtu cd7 ((b)(4)). It was noted that staining with bond rtu cd7 was giving a different staining pattern to the concentrate version i. E. (b)(4).
Patient Sequence No: 1, Text Type: D, B5


[9630541] Internal investigation revealed that cd2 cells were grown as cd7 and likewise, cd7 cells as cd2. Both cd2 and cd7 are membrane markers of most t-cells and nk-cells and are used as part of a panel alongside other t-cell markers. Since cd2 and cd7 expression may vary between reactive t-cell proliferations and t-cell lymphomas, the decision was made to conduct a recall. Pa0271 lots 08173, 11855, 14784, pa0266 lots 10604, 13566, ncl-cd7-580 lots l158018, l158019, l158020. Our course of action is to release a field correction letter to inform all customers of the affected lots of this issue and request any unused product to be returned. Clinicians are also advised that they may wish to review results obtained from these lots. Corrective actions include review of processes in tissue culture production and implement tightened line-clearance procedure to avoid further mix-ups.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3004859032-2012-00001
MDR Report Key2455347
Report Source07
Date Received2012-02-06
Date of Report2011-12-06
Date of Event2011-12-05
Date Added to Maude2012-08-14
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer StreetBALLIOL BUSINESS PARK BENTON LANE
Manufacturer CityNEWCASTLE UPON TYNE NE128EW
Manufacturer CountryUK
Manufacturer PostalNE12 8EW
Manufacturer Phone912150567
Single Use3
Remedial ActionOT
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameBOND AND NOVOCASTRA
Generic NameIMMUNOHISTOCHEMISTRY REAGEANT
Product CodeNJT
Date Received2012-02-06
Model NumberPA0271,PA0266, CD7
Catalog NumberPA0271, PA0266, CD7
Lot NumberVARIOUS
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerLEICA BIOSYSTEMS NEWCASTLE LTD
Manufacturer AddressBALLIOL BUSINESS PARK BENTON LANE NEWCASTLE UPON TYNE NE128EW UK NE12 8EW


Patients

Patient NumberTreatmentOutcomeDate
10 2012-02-06

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.