OLYMPUS URETERO-RENO VIDEOSCOPE URF-V UNKNOWN

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2012-01-31 for OLYMPUS URETERO-RENO VIDEOSCOPE URF-V UNKNOWN manufactured by Olympus Medical Systems Corporation.

Event Text Entries

[15861865] Olympus was informed that "rubber on tip damaged by laser". Olympus followed up with the user facility and was informed that this device had been used in an unspecified urology procedure. There was no pt harm reported.
Patient Sequence No: 1, Text Type: D, B5


[15942964] Olympus followed up with the user facility to obtain additional information regarding this report. The user facility stated that the reported phenomenon could have likely occurred during an urology procedure and there was no pt injury reported. There was no additional information provided. The device referenced in this report was returned to olympus for evaluation. The evaluation confirmed the user's report. The mesh unit at the lower portion of the bending section was damaged which caused a tear on the bending cover, and a frayed wire was found protruding from the bending cover, causing a leaks. The damage mesh unit was attributed to physical damage due to user handling. This report is being submitted as an mdr in an abundance of caution.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number8010047-2012-00023
MDR Report Key2455475
Report Source06
Date Received2012-01-31
Date of Report2012-01-05
Date Mfgr Received2012-01-05
Date Added to Maude2012-04-19
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactLAURA STORMS-TYLER
Manufacturer Street2400 RINGWOOD AVENUE
Manufacturer CitySAN JOSE CA 95131
Manufacturer CountryUS
Manufacturer Postal95131
Manufacturer Phone4848965688
Manufacturer G1OLYMPUS MEDICAL SYSTEMS CORPORATION
Manufacturer Street2951 ISHIKAWA-CHO
Manufacturer CityHACHIOJI-SHI, TOKYO 192-8507
Manufacturer CountryJA
Manufacturer Postal Code192-8507
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameOLYMPUS URETERO-RENO VIDEOSCOPE
Generic NameURETEROSCOPE
Product CodeGCQ
Date Received2012-01-31
Returned To Mfg2012-01-04
Model NumberURF-V
Catalog NumberUNKNOWN
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerOLYMPUS MEDICAL SYSTEMS CORPORATION
Manufacturer Address2951 ISHIKAWA-CHO HACHIOJI-SHI, TOKYO 192-8507 JA 192-8507


Patients

Patient NumberTreatmentOutcomeDate
10 2012-01-31

© 2026 FDA.report