IMMAGE IMMUNOCHEMISTRY SYSTEM RHEUMATOID FACTOR (RF) REAGENT 447070

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2012-02-16 for IMMAGE IMMUNOCHEMISTRY SYSTEM RHEUMATOID FACTOR (RF) REAGENT 447070 manufactured by Beckman Coulter Inc..

Event Text Entries

[2537682] Customer called on (b)(4) 2012, reporting of false positive (>20 iu/ml) immage immunochemistry system rheumatoid factor (rf) reagent results generated on the immage 800 immunochemistry system. Customer stated that rf reagent lot m106133 was the cause of the false positive results and rerunning the sample will usually give the correct results. Customer noted this has been a problem on and off since early december when they first started using lot m106133 and believes that the issue is with the reagent and not the instrument. Customer stated that no erroneous results have been reported outside the lab and therefore no change to patient treatment was attributed to this event. Customer did not provide any patient test results or information related to this event. No service was generated or requested for this event as customer believes that this was a reagent related issue.
Patient Sequence No: 1, Text Type: D, B5


[9714773] Evaluation code - results code - (other): root cause is unknown but appears to be a reagent issue.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2050012-2012-00494
MDR Report Key2457082
Report Source05,06
Date Received2012-02-16
Date of Report2012-01-31
Date of Event2012-01-31
Date Mfgr Received2012-01-31
Device Manufacturer Date2011-07-19
Date Added to Maude2012-08-17
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS. DUNG NGUYEN
Manufacturer Street250 S. KRAEMER BOULEVARD
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal92821
Manufacturer Phone7149614941
Manufacturer G1BECKMAN COULTER INC.
Manufacturer Street250 S. KRAEMER BLVD
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal Code92821
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameIMMAGE IMMUNOCHEMISTRY SYSTEM RHEUMATOID FACTOR (RF) REAGENT
Generic NameSYSTEM, TEST, RHEUMATOID FACTOR
Product CodeDHR
Date Received2012-02-16
Catalog Number447070
Lot NumberM106133
Device Expiration Date2013-07-31
OperatorHEALTH PROFESSIONAL
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerBECKMAN COULTER INC.
Manufacturer Address250 S. KRAEMER BLVD BREA CA 92821 US 92821


Patients

Patient NumberTreatmentOutcomeDate
10 2012-02-16

© 2026 FDA.report