CHAMPION 54P UNKNOWN

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2012-02-23 for CHAMPION 54P UNKNOWN manufactured by Champion Mfg Inc..

Event Text Entries

[2488828] A brand new electric cardiac chair was just introduced into the hospital for our post-surgical cardiac patients. The patient was placed from the bed to the new chair. The cardiac monitor displayed severe 60 cycle interference, thus making it impossible to properly display the ekg tracing on the physiological monitor. The patient was removed from the chair and the 60 cycle ac (alternate current) interference disappeared. ======================manufacturer response for 54p electric cardiac chair, champion (per site reporter)======================we spoke with an engineer at champion mfg, inc. And he said that this device didn't have an emi (electromagnetic interference) certificate. They are aware of the problem of the monitor interference. He responded to us they have no intention of attempting to correct this issue.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2473541
MDR Report Key2473541
Date Received2012-02-23
Date of Report2012-02-22
Date of Event2012-02-19
Report Date2012-02-22
Date Reported to FDA2012-02-23
Date Added to Maude2012-03-01
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameCHAMPION
Generic Name54P ELECTRIC CARDIAC CHAIR
Product CodeGBB
Date Received2012-02-23
Model Number54P
Catalog NumberUNKNOWN
Lot Number*
ID Number*
OperatorNURSE
Device AvailabilityY
Device Age1 DY
Device Sequence No1
Device Event Key0
ManufacturerCHAMPION MFG INC.
Manufacturer AddressONE INVACARE WAY ELYRIA OH 44035 US 44035


Patients

Patient NumberTreatmentOutcomeDate
10 2012-02-23