HUT EXT DR FINAL ASSY-REVERSE 404007

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06,07 report with the FDA on 2012-02-28 for HUT EXT DR FINAL ASSY-REVERSE 404007 manufactured by Liebel-flarsheim Co.

Event Text Entries

[2474555] (b)(6): customer reports when staff attempted to start the urology system for days procedures, the monitors failed to work. Patient procedures were canceled. Customer does not have a back up room. No reported injuries. Customer reports all procedures have been delayed pending room repair.
Patient Sequence No: 1, Text Type: D, B5


[9610293] Field service engineer (fse) troubleshot the no fluoro image problem and found both monitors on tower worked in last image hold (lih) channel 11. Fse then swapped video input at the monitors and image swapped, then fse went to bnc on platinum one computer and flexed it and the monitor came up in channel 10. Fse determined from this that the unit needs a bnc cable ph 704051 or video pcb pn 710293. The unit is currently working. When fse asked for approval from customer for parts replacement, customer wanted to open another service request with different po# to cover video pcb 710293, and bnc cable 704051 and labor at a later date. Fse completed hydravision service checklist with reference service manual, and found all readings within tolerance. Unit passed checkout procedures and was returned to full service by the customer.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1518293-2012-00041
MDR Report Key2480328
Report Source05,06,07
Date Received2012-02-28
Date of Report2012-02-28
Date of Event2012-02-16
Date Mfgr Received2012-02-16
Device Manufacturer Date2005-07-01
Date Added to Maude2012-08-24
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactDAVID BENSON
Manufacturer Street2111 EAST GALBRAITH RD.
Manufacturer CityCINCINNATI OH 45237
Manufacturer CountryUS
Manufacturer Postal45237
Manufacturer Phone5139485719
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameHUT EXT DR FINAL ASSY-REVERSE
Product CodeKQS
Date Received2012-02-28
Model NumberHUT EXT DR
Catalog Number404007
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerLIEBEL-FLARSHEIM CO
Manufacturer Address2111 EAST GALBRAITH RD. CINCINNATI OH 45237 US 45237


Patients

Patient NumberTreatmentOutcomeDate
10 2012-02-28

© 2026 FDA.report