INDEPENDENCE IBOT 4000 MOBILITY SYSTEM IT004011

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 04 report with the FDA on 2012-03-01 for INDEPENDENCE IBOT 4000 MOBILITY SYSTEM IT004011 manufactured by Independence Technology, L.l.c..

Event Text Entries

[2552203] User reported a forward fall of the device while attempting to descend a set of exterior cement stairs in stair function. No injury was reported. User states that he had a previous injury (pulled muscle in arm) and while holding onto the railing to descend steps, because of his arm injury, he was unable to lean back properly/enough to keep the device from rotating its wheel cluster, and the chair moved out from underneath him and proceeded to descend the stairs on its own. The device fell to the ground, while user was still holding onto the railing. User states that the device sustained no damage, other than a scuff mark on the back of the armrest. User was not injured, and he had assistance getting back into and using the device. User was advised by service hotline that the event may have occurred quickly and not allowed proper exit from cluster safety lock, but also that the driver/assistant should have control of the device while in stair function. It was suggested to the user to perhaps have assistance with stair climbing while injured. Field service was dispatched to inspect the device, clear the service wrench and retrieve the electronic configuration file (ecf) for review. (b)(4).
Patient Sequence No: 1, Text Type: D, B5


[9758953] Service was dispatched to inspect the device, retrieve the electronic configuration file (ecf) for evaluation, and clear the service wrench. A field service activity/device checkout report (esar) was forwarded to the complaint handling unit (chu) per standard operating procedure. The engineering review of the ecf determined that on (b)(6) 2012 at 13:45:45 gmt, the device entered stair function and recorded a frame lean stop. Twenty six (26) seconds later, the device recorded a controller alert indicating a cluster safety lock condition while descending. At the same time stamp, the device went to a controller failure condition because of exceeding its pitch limit. No other alarms were present in the logs that would have contributed to this event. The black box data for the event shows that a one step descent was successfully completed with proper control. A second step is descended with inadequate pitch correction by the user, allowing for a rapid forward pitch of the device and cluster safety lock condition. The device then continued to pitch forward until the limit w as exceeded. The conclusion is that the device did not malfunction. The device entered cluster safety lock, and went to controller failure due to exceeding its pitch limit. Cluster safety lock is the devices detection of a loss of control during stair climbing and is consistent with the user's report of the circumstances of the event.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3003508375-2012-00001
MDR Report Key2483433
Report Source04
Date Received2012-03-01
Date of Report2012-03-01
Date of Event2012-02-25
Date Mfgr Received2012-02-27
Device Manufacturer Date2006-06-01
Date Added to Maude2012-08-01
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMICHAEL O'MEARA, DIRECTOR
Manufacturer StreetROUTE 22 WEST
Manufacturer CitySOMERVILLE NJ 08876
Manufacturer CountryUS
Manufacturer Postal08876
Manufacturer Phone9087223767
Manufacturer G1CREATIVE TECHNOLOGY SERVICES
Manufacturer Street7444 HAGGERTY ROAD
Manufacturer CityCANTON MI 48187
Manufacturer CountryUS
Manufacturer Postal Code48187
Single Use3
Remedial ActionIN
Previous Use Code3
Removal Correction NumberNOT APPLICABLE
Event Type3
Type of Report3

Device Details

Brand NameINDEPENDENCE IBOT 4000 MOBILITY SYSTEM
Generic NameSTAIR CLIMBING WHEELCHAIR
Product CodeIMK
Date Received2012-03-01
Model NumberIBOT
Catalog NumberIT004011
Lot NumberNOT APPLICABLE
ID NumberNOT APPLICABLE
OperatorLAY USER/PATIENT
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerINDEPENDENCE TECHNOLOGY, L.L.C.
Manufacturer AddressSOMERVILLE NJ 08876 US 08876


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2012-03-01

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.