INGENUITY TF PET/MR 882380

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 07 report with the FDA on 2012-02-28 for INGENUITY TF PET/MR 882380 manufactured by Philips Medical Systems(cleveland), Inc..

Event Text Entries

[2548512] Philips received a customer complaint stating a photopenic artifact was detected through the lower lungs on the magnetic resonance attenuation corrected (mrac) image. The customer also noticed low uptake in the inferior wall of the heart that was also detected on the non-attenuation corrective (nac) images. The artifact is not present on the mr map or nac dataset.
Patient Sequence No: 1, Text Type: D, B5


[9670837] (b)(4). There is a remote potential for false-negative interpretation in pts with neoplasm that could result in lack of treatment leading to the necessity for medical intervention to prevent permanent impairment. There was no report of harm to a pt in this case and the harm is not expected to occur due to trained interpreting physician recognizing image artifact, and default display of non-ac and attenuation map for comparison. Instructions for use includes review of non-attenuation correction images (459800153451 rev. B p18). This reported event is currently under investigation. A follow-up report will be sent when the investigation is completed.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1525965-2012-00004
MDR Report Key2483586
Report Source07
Date Received2012-02-28
Date of Report2012-02-03
Date of Event2012-01-24
Date Mfgr Received2012-02-03
Date Added to Maude2012-08-23
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactKUMUDINI CARTER
Manufacturer Street595 MINER RD.
Manufacturer CityCLEVELAND OH 44143
Manufacturer CountryUS
Manufacturer Postal44143
Manufacturer Phone4404833032
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameINGENUITY TF PET/MR
Generic NameINGENUITY TF PET/MR
Product CodeOUO
Date Received2012-02-28
Model Number882380
Catalog Number882380
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerPHILIPS MEDICAL SYSTEMS(CLEVELAND), INC.
Manufacturer Address595 MINER RD. CLEVELAND OH 44143 US 44143


Patients

Patient NumberTreatmentOutcomeDate
10 2012-02-28

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.