BOND AND NOVOCASTRA NCL-ER-6F11

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 04 report with the FDA on 2012-03-09 for BOND AND NOVOCASTRA NCL-ER-6F11 manufactured by Leica Biosystems Newcastle Ltd.

Event Text Entries

[2404129] Customer raised query ((b)(4)) regarding potential false positive results obtained using leica estrogen receptor clone ((b)(4)). Discordant results were reported on a retrospective validation cohort when compared with staining results obtained with competitor er antibody. The same staining pattern was subsequently observed in pt samples and the hospital has suspended er testing service resulting in a potential delay in pt diagnosis.
Patient Sequence No: 1, Text Type: D, B5


[9696676] Review of the slides from (b)(6) hospital and subsequent assessment at (b)(6) has confirmed that the protocol at (b)(6) hospital, which employed a more aggressive pretreatment, is the most likely cause of staining pattern noted at by the customer. When an appropriate epitope retrieval buffer was employed the staining was not observed. Advised customer to revalidate the antibody using the appropriate epitope retrieval solution. The complaint is still under investigation. A follow-up report will be submitted after we complete our investigations.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3004859032-2012-00002
MDR Report Key2492795
Report Source04
Date Received2012-03-09
Date of Report2012-03-08
Date of Event2012-02-01
Date Mfgr Received2012-02-01
Device Manufacturer Date2012-01-11
Date Added to Maude2012-08-28
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer StreetBENTON LANE BALLIOL BUSINESS PARK
Manufacturer CityNEWCASTLE UPON TYNE NE128EW
Manufacturer CountryUK
Manufacturer PostalNE12 8EW
Manufacturer Phone912150567
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameBOND AND NOVOCASTRA
Generic NameIMMUNOHISTOCHEMISTRY REAGENT
Product CodeMYA
Date Received2012-03-09
Model NumberNCL-ER-6F11
Catalog NumberNCL-ER-6F11
Lot Number6008295
Device Expiration Date2014-01-31
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerLEICA BIOSYSTEMS NEWCASTLE LTD
Manufacturer AddressBENTON LANE BALLIOL BUSINESS PARK NEWCASTLE UPON TYNE NE128EW UK NE12 8EW


Patients

Patient NumberTreatmentOutcomeDate
10 2012-03-09

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.