KODAK EKTASCAN LASER PRINTER 2180 1259965

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 1999-11-09 for KODAK EKTASCAN LASER PRINTER 2180 1259965 manufactured by Eastman Kodak Co. Health Imaging Division.

Event Text Entries

[20136797] The latching mechanism of the film supply magazine has a mechanism that guards against the lid of the film magazine falling down unexpectedly on someone's hand during the loading of film. A cap screw in this mechanism was loose and mechanism failed. The result was the lid fell on the technician's hand while she was loading the film. This resulted in a fracture to her hand.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1315356-1999-00001
MDR Report Key249561
Report Source05,06
Date Received1999-11-09
Date of Report1999-11-08
Date of Event1999-10-08
Date Mfgr Received1999-10-08
Date Added to Maude1999-11-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer ContactJUDITH WALLACE
Manufacturer Street91 ELMGROVE RD
Manufacturer CityROCHESTER NY 14653
Manufacturer CountryUS
Manufacturer Postal14653
Manufacturer Phone7167246111
Manufacturer G1*
Manufacturer Street*
Manufacturer City*
Manufacturer Country*
Single Use3
Remedial ActionRL
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameKODAK EKTASCAN LASER PRINTER 2180
Generic NameX-RAY MEDICAL IMAGER-
Product CodeLMC
Date Received1999-11-09
Model Number2180
Catalog Number1259965
Lot Number*
ID Number*
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Implant FlagN
Date RemovedA
Device Sequence No1
Device Event Key241780
ManufacturerEASTMAN KODAK CO. HEALTH IMAGING DIVISION
Manufacturer Address343 STATE ST. ROCHESTER NY 14650 US
Baseline Brand NameKODAK EKTASCAN LASER PRINTER 2180
Baseline Generic NameX-RAY MEDICA IMAGER
Baseline Model No2180
Baseline Catalog No1259965
Baseline ID*
Baseline Device FamilyEKTASCAN LASER PRINTER
Baseline Shelf Life [Months]NA
Baseline PMA FlagN
Baseline 510K PMNY
Premarket NotificationK921743
Baseline PreamendmentN
Baseline TransitionalN
510k ExemptN


Patients

Patient NumberTreatmentOutcomeDate
10 1999-11-09