HU-FRIEDY EXPRO XPC44757C

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2012-03-16 for HU-FRIEDY EXPRO XPC44757C manufactured by .

Event Text Entries

[2554014] During the course of a root canal procedure, a doctor was using a hu-friedy probe to measure a canal abscess. The following week, the pt returned to the doctor for a f/u and the doctor noticed that a small piece of the tip of the probe was embedded in the pt's tissue. The doctor informed the pt and indicated that he would need to make a small surgical incision to remove the embedded tip. The doctor proceeded to remove the tip and the surgical site healed w/o any further complications.
Patient Sequence No: 1, Text Type: D, B5


[9769137] The returned instrument is broken near the 3 mm marking. Fracture surface shows no signs of corrosion. Both working end s of the returned instrument are bent and do not match hu-friedy master sample in shape. Returned expo was visually inspected under magnification (20x) and the remaining section of the probe end and the explorer working end were found to have deep nicks and scratches. The deep nicks and scratches are indication of the tips getting caught in the holes of an ultrasonic cleaner basket/tray over time. This causes tips to bend and if working end are re-shaped, this type of movement can lead to low cycle fatigue whereby a force higher than material capability can lead to breakage of the tip. Additionally, these types of instruments (xpc44757c) have a normal life expectancy of up to 4 yrs. The returned instrument was manufactured over 5 yrs ago (aug. 2006).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1416605-2012-00001
MDR Report Key2496662
Report Source06
Date Received2012-03-16
Date of Report2012-03-09
Date Mfgr Received2012-02-23
Device Manufacturer Date2006-08-01
Date Added to Maude2012-08-02
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationDENTIST
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMARIA VRABIE, TEAM LEADER OF RA
Manufacturer Street3232 N ROCKWELL STREET
Manufacturer CityCHICAGO IL 60618
Manufacturer CountryUS
Manufacturer Postal60618
Manufacturer Phone7739753495
Single Use3
Remedial ActionOT
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameHU-FRIEDY EXPRO
Generic NameXPC44757C
Product CodeEIX
Date Received2012-03-16
Returned To Mfg2012-02-23
Model NumberXPC44757C
Catalog NumberXPC44757C
Lot NumberV6
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeYR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2012-03-16

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.