BURDICK ECLIPSE PREMIER E10-6AG17 NA

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,05 report with the FDA on 2012-03-15 for BURDICK ECLIPSE PREMIER E10-6AG17 NA manufactured by Cardiac Science Corporation.

Event Text Entries

[2629262] The hospital's cardio-respiratory services mgr who reported the incident to cardiac science said one of their ecg machines had a "circuit board fire. " the hospital's clinical engineer dismantled the machine and found there were 2 surface mount components close to the power inlet that were charred. There was no pt injury.
Patient Sequence No: 1, Text Type: D, B5


[9764557] The hospital's clinical engineer found that there were 2 surface mount components that were charred and a picture of the board with the damaged components was sent to cardiac science. These components meet iec 60601 requirements and may have caused smoke when they failed, but would not have caused a fire. The failure resulted in the inability of the device to meet its essential function, but there was no hazard to the pt. At the present time, the facility does not intend to return the device or board to cardiac science for eval.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3014398-2012-00002
MDR Report Key2498824
Report Source01,05
Date Received2012-03-15
Date of Report2012-02-24
Date of Event2012-02-24
Date Mfgr Received2012-02-24
Device Manufacturer Date2008-07-01
Date Added to Maude2012-03-21
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactBEV MAGRANE
Manufacturer Street3303 MONTE VILLA PKWY.
Manufacturer CityBOTHELL WA 98021
Manufacturer CountryUS
Manufacturer Postal98021
Manufacturer Phone4254022365
Manufacturer G1CARDIAC SCIENCE CORPORATION
Manufacturer Street500 BURDICK PKWY
Manufacturer CityDEERFIELD WI 53531969
Manufacturer CountryUS
Manufacturer Postal Code53531 9692
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameBURDICK ECLIPSE PREMIER
Generic NameRESTING ECG SYSTEM
Product CodeLOS
Date Received2012-03-15
Model NumberE10-6AG17
Catalog NumberNA
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerCARDIAC SCIENCE CORPORATION
Manufacturer AddressBOTHELL WA US


Patients

Patient NumberTreatmentOutcomeDate
10 2012-03-15

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.