CONMED DISPOSABLE MICROBIOLOGY BRUSH 130

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2012-03-21 for CONMED DISPOSABLE MICROBIOLOGY BRUSH 130 manufactured by Conmed Corporation.

Event Text Entries

[2632537] It was reported, "performed a lavage on patient. Cytology brushes were then used and tiny blue pieces appeared in the patient's airway. " it also was reported that no patient injury resulted from the event.
Patient Sequence No: 1, Text Type: D, B5


[9770913] This is being reported to the fda due to the nature of the complaint: unretrieved device fragments (udf). When the quality engineering evaluation is completed a supplemental report will be filed.
Patient Sequence No: 1, Text Type: N, H10


[9799567] The suspect device is a conmed disposable, single-use microbiology brush designed to provide a method of obtaining contamination free specimens, primarily from the lower respiratory tract. It consists of a catheter within a catheter with a centrally located sampling brush. A wax-like plug prevents contamination of the inner catheter and brush before sampling. The plug will dissolve safely after being expelled from the catheter. This device is indicated to extract sterile microbiological specimens of the bronchial tree without contamination from the upper airway or bronchoscope. A review of the manufacturing documents from the dhr/lhr, device history record/lot history record, for lot 1111284 has verified the devices were produced according to current and approved procedures and material specifications. Non-conformances regarding the product's identity, quality, safety, effectiveness or performance were not identified during manufacture. Two (2) used devices were returned to the conmed complaint handling center for investigation. The returned device were examined in the laboratory. Both devices were visually checked for any damage. No scratches or kinks or damage of any kind was observed on each device that could account for any device fragments separating from the device. Both devices were observed as undamaged, and, intact. There was no conclusive evidence of any product damage noticed during the examination that could cause the customer reported complaint failure. This incident occurred when a conmed disposable, single-use microbiology brush was used in conjunction with a fairly new boston scientific endoscope. From conversations with hospital endoscopy staff, it is now believed that the udf's, unretrieved device fragments, are from the boston scientific scope. This stems from the fact that the same "tiny blue pieces" that appeared in the patient's airway have also occurred in numerous other procedures where the boston scientific scope is the common factor. Many of the procedures that the same blue udf's have occurred where a conmed device was never utilized in the procedure, yet, the boston scientific endoscope was utilized. The complaint investigation has not identified nor confirmed any manufacturing defects associated with the returned devices. No corrective action is recommended at this time. Conmed is considering this complaint closed.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3007305485-2012-00020
MDR Report Key2498927
Report Source06
Date Received2012-03-21
Date of Report2012-04-16
Date of Event2012-02-14
Date Mfgr Received2012-02-21
Device Manufacturer Date2011-11-28
Date Added to Maude2012-03-23
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMR. STEPHEN CASANOVA
Manufacturer Street525 FRENCH ROAD
Manufacturer CityUTICA NY 135025994
Manufacturer CountryUS
Manufacturer Postal135025994
Manufacturer Phone3156243463
Manufacturer G1CONSOLIDATED MEDICAL EQUIPMENT COMPANY
Manufacturer StreetAVE. ALEJANDRO DUMAS NO. 11321 COMPLEJO INDUSTRIAL CHIHUAHUA
Manufacturer CityCHIHUAHUA, CHIHUAHUA CP 1136
Manufacturer CountryMX
Manufacturer Postal CodeCP 1136
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCONMED DISPOSABLE MICROBIOLOGY BRUSH
Generic NameMICROBIOLOGY BRUSH
Product CodeBTG
Date Received2012-03-21
Returned To Mfg2012-03-02
Catalog Number130
Lot Number1111284
Device Expiration Date2016-11-01
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerCONMED CORPORATION