BELIMED CS720

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,07 report with the FDA on 2012-01-12 for BELIMED CS720 manufactured by Belimed Gmbh.

Event Text Entries

[15859915] A sterile processing department operator was using a belimed cs720 cart washer. The doors for the cart washer are closed by pressing and holding two buttons simultaneously the entire time the doors are closing. When the operator pressed and held these two buttons, a cart got stuck in between the two doors. The operator then pressed and held the two buttons with one hand while pushing the cart into the wash chamber with the other hand. The doors then closed on the operator's arm causing an injury. The operator's manual states the following: "if you must be inside the wash chamber for any reason, press the emergency stop push-button inside the chamber. This will prevent any operation of the washer. "
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number8043640-2012-00001
MDR Report Key2501431
Report Source05,07
Date Received2012-01-12
Date of Report2012-01-04
Date of Event2011-12-14
Date Facility Aware2011-12-14
Report Date2012-01-04
Date Reported to Mfgr2011-12-15
Date Mfgr Received2011-12-15
Date Added to Maude2012-03-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional0
Initial Report to FDA3
Report to FDA0
Event Location3
Manufacturer Street2284 CLEMENTS FERRY ROAD
Manufacturer CityCHARLESTON SC 29492
Manufacturer CountryUS
Manufacturer Postal29492
Manufacturer G1BELIMED DEUTSCHLAND GMBH
Manufacturer StreetEDISONSTRASSE 7-A
Manufacturer CityMUEHLDORF AM INN 84453
Manufacturer CountryGM
Manufacturer Postal Code84453
Single Use0
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameBELIMED
Generic NameWASHER OF BODY WASTE RECEPTACLES
Product CodeFLH
Date Received2012-01-12
Model NumberCS720
OperatorHEALTH PROFESSIONAL
Device AgeDA
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerBELIMED GMBH
Manufacturer AddressMUEHLDORF GM


Patients

Patient NumberTreatmentOutcomeDate
101. Hospitalization 2012-01-12

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.