RAPID RESPONSE HUT TABLE 1217

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 04,05,06,07 report with the FDA on 2012-03-02 for RAPID RESPONSE HUT TABLE 1217 manufactured by Medical Positioning, Inc..

Event Text Entries

[2574922] As the nurse plugged the product into 120 vac utility power, there was a momentary spark between the outlet and the plug and the nurse received a minor shock, but no injury from the spark.
Patient Sequence No: 1, Text Type: D, B5


[9748170] Investigation uncovered that the user was attempting to plug the product into the 120 vac utility power while the power switch is in the "on" position. Because there is a transformer in the unit, there is a momentary spike in the current draw and then levels out. The company advised the customer to make sure that the product's power switch is in the "off" position before connecting it to utility power. The company also opened capa (b)(4). Upon review of product labeling it was discovered that instructions for properly connecting the product to utility power are missing from the owner's manual. The company will update the manual with the proper instructions/warning for connecting the unit to utility power. The company will send the manual update to all nine (9) customers who have purchased the product.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1932056-2012-00001
MDR Report Key2507465
Report Source04,05,06,07
Date Received2012-03-02
Date of Report2012-02-29
Date of Event2012-02-28
Date Mfgr Received2012-02-28
Device Manufacturer Date2012-01-25
Date Added to Maude2012-07-31
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactJOE HURTIG, QUALITY MANAGER
Manufacturer Street1717 WASHINGTON
Manufacturer CityKANSAS CITY MO 64108
Manufacturer CountryUS
Manufacturer Postal64108
Manufacturer Phone8165952356
Single Use3
Remedial ActionRB
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameRAPID RESPONSE HUT TABLE
Generic NameRAPID RESPONSE HUT TABLE
Product CodeLGX
Date Received2012-03-02
Model Number1217
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMEDICAL POSITIONING, INC.
Manufacturer AddressKANSAS CITY MO US


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2012-03-02

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.