GCE COMBILITE I VIPR, 40 MICRONS TYPE 0710092 FOR OXYGEN

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,06 report with the FDA on 2011-02-28 for GCE COMBILITE I VIPR, 40 MICRONS TYPE 0710092 FOR OXYGEN manufactured by .

Event Text Entries

[2574474] On (b)(6) 2010, linde was initially informed that an incident involving a gce combilite i vipr, 40 microns type (b)(4) for oxygen had occurred in the (b)(6). No person was injured. On (b)(6) 2011, (b)(6) communicated that although a root cause was not identified, similar mechanism for ignition could not be excluded in gce valves of similar design. Therefore (b)(6) 2011, is regarded as the linde awareness date for regulatory reporting of the present incident and despite the fact that this specific valve is not sold in the us the case will be reported to fda. It was reported by (b)(6), that a medical oxygen cylinder had ignited when one of their firemen was performing routine testing of the cylinder. The cylinder had been removed from its bag and visually inspected. On opening the cylinder valve, the operator reported seeing a flame coming from the valve outlet. He then dropped the cylinder, the fire was quickly extinguished using a foam extinguisher.
Patient Sequence No: 1, Text Type: D, B5


[9838503] (b)(4). Accidental ignition/fire in oxygen valve, not pt/personnel harm. Cause for ignition not known.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3003900188-2011-00031
MDR Report Key2507737
Report Source01,06
Date Received2011-02-28
Date Mfgr Received2011-01-31
Device Manufacturer Date2008-08-01
Date Added to Maude2012-04-13
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional0
Initial Report to FDA0
Report to FDA0
Event Location0
Manufacturer CountrySW
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameGCE COMBILITE I VIPR, 40 MICRONS TYPE 0710092 FOR OXYGEN
Generic NameNONE
Product CodeECX
Date Received2011-02-28
Model Number0710092
Lot Number4280965
OperatorOTHER
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0


Patients

Patient NumberTreatmentOutcomeDate
10 2011-02-28