TMAX & DISPOSABLE FACE MASK FOR POSITIONING 74-00-04 R1

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 00,06 report with the FDA on 2012-02-14 for TMAX & DISPOSABLE FACE MASK FOR POSITIONING 74-00-04 R1 manufactured by Tenet Medical Engineering.

Event Text Entries

[2577335] A pressure ulcer developed on a pt's chin area under the "disposable face mask for positioning" during a beach chair shoulder surgery, prolonged (7hrs) pressure was delivered to the pts chin from a face mask intended to stabilize the pt's head during surgery. In discussions with the health care provider they indicated that the hosp may not have applied the leg belt in the appropriate position, and that during the unusually long procedure the pt may have shifted slightly causing additional pressure to the area. Initial discussions with the health care provider indicated that the pt would recover from the injury without medical or surgical intervention.
Patient Sequence No: 1, Text Type: D, B5


[9724027] Initial reports from the health care provider indicated that the injury to the pt was minor as healing was expected to occur without medical or surgical intervention. The health care provider and the mfr agree that the cause of the injury was due to the exceptionally long duration of the procedure in conjunction with the mask potentially being too tight at the beginning of the surgery or becoming too tight as a result of pt movement during the procedure. It was discovered that the pt leg restraint was not being used appropriately and may have contributed to pt movement if movement occurred.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9681741-2011-00001
MDR Report Key2508474
Report Source00,06
Date Received2012-02-14
Date of Report2011-02-11
Date of Event2011-01-07
Date Mfgr Received2011-01-14
Date Added to Maude2012-07-30
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactBRENT KING
Manufacturer Street#203 - 11979 40TH ST. S.E.
Manufacturer CityCALGARY, AB
Manufacturer CountryCA
Manufacturer Phone5710750
Single Use3
Remedial ActionPM
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTMAX & DISPOSABLE FACE MASK FOR POSITIONING
Generic NameBEACH CHAIR AND FACE MASK
Product CodeBWN
Date Received2012-02-14
Model Number74-00-04 R1
Lot NumberNOT PROVIDED
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerTENET MEDICAL ENGINEERING
Manufacturer Address#203 - 11979 40TH ST. S.E. CALGARY, AB T2Z4M3 CA T2Z4M3


Patients

Patient NumberTreatmentOutcomeDate
10 2012-02-14

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.