DRAGONFLY IMAGING CATHETER 13751-02 NA

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06,07 report with the FDA on 2012-03-27 for DRAGONFLY IMAGING CATHETER 13751-02 NA manufactured by St Jude Medical.

Event Text Entries

[2596692] In preparation for a case with oct imaging of a stent in the mid lad, a dissection of the very short left-main was caused by the guiding catheter. This was not apparent until after oct imaging. Very poor blood clearance was obtained such that the physician assumed that the guiding catheter had slipped out prior to the pullback. After the pullback, the pt went into ventricular fibrillation and a large dissection extending along the lad and possibly back to the aorta. Pt was stabilized and left-main and lad were successfully treated with a new proximal stent. Physician stated that absolutely there was no suggestion that the imaging catheter caused this incident and he had a similar experience recently (in a non oct imaging case).
Patient Sequence No: 1, Text Type: D, B5


[9705884] Based on the info provided incident was not caused by the dragonfly imaging catheter used for oct imaging. No product was returned. Review of the device history record was not possible since the lot number was unavailable.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3004672267-2012-00002
MDR Report Key2508840
Report Source05,06,07
Date Received2012-03-27
Date of Report2012-03-22
Date of Event2012-02-29
Date Mfgr Received2012-02-29
Date Added to Maude2012-03-29
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactANN GRAVES
Manufacturer StreetONE TECHNOLOGY PARK DRIVE
Manufacturer CityWESTFORD MA 01886
Manufacturer CountryUS
Manufacturer Postal01886
Manufacturer Phone9786921408
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameDRAGONFLY IMAGING CATHETER
Generic NameINTERVASCULAR IMAGING CATHETER
Product CodeNQQ
Date Received2012-03-27
Model Number13751-02
Catalog NumberNA
Lot NumberUNK
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerST JUDE MEDICAL
Manufacturer AddressWESTFORD MA US


Patients

Patient NumberTreatmentOutcomeDate
101. Hospitalization; 2. Required No Informationntervention 2012-03-27

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.