VITROS 5,1 FS CHEMISTRY SYSTEM 6801375

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 00,01,05 report with the FDA on 2012-03-29 for VITROS 5,1 FS CHEMISTRY SYSTEM 6801375 manufactured by Ortho-clinical Diagnostics.

Event Text Entries

[2597625] The customer obtained multiple, imprecise vitros phyt quality control results (results = 140. 2, , 149. 7, 151. 9, 133. 1, 146. 9 g/ml vs. Expected result = 109. 0 g/ml) while using the vitros 5,1 fs chemistry system. Biased results of the direction and magnitude observed may lead to inappropriate physician action if the events were to recur with patient samples. There was no report of affected patient samples. There was no allegation of harm to patients as a result of this event. This report corresponds to ortho clinical diagnostics, inc. (b)(4).
Patient Sequence No: 1, Text Type: D, B5


[9706787] The investigation confirmed that multiple, imprecise vitros phyt quality control results were obtained while using the vitros 5,1 fs analyzer. Patient samples were not reported to have been affected. An ocd field engineer performed service actions to multiple analyzer subsystems and performed related adjustments. Performance testing following service could not confirm that the equipment was returned to expected operation. There is no evidence that the vitros phyt reagent slides malfunctioned. The investigation is ongoing. The assignable root cause is an instrument related issue.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1319681-2012-00049
MDR Report Key2509291
Report Source00,01,05
Date Received2012-03-29
Date of Report2012-03-29
Date of Event2012-02-11
Date Mfgr Received2012-03-01
Device Manufacturer Date2010-10-02
Date Added to Maude2012-04-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMR. JOSEPH FALVO
Manufacturer Street100 INDIGO CREEK DRIVE
Manufacturer CityROCHESTER NY 14626
Manufacturer CountryUS
Manufacturer Postal14626
Manufacturer Phone5854533000
Manufacturer G1ORTHO-CLINICAL DIAGNOSTICS
Manufacturer Street100 INDIGO CREEK DRIVE
Manufacturer CityROCHESTER NY 14626
Manufacturer CountryUS
Manufacturer Postal Code14626
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameVITROS 5,1 FS CHEMISTRY SYSTEM
Generic NameCHEMISTRY ANALYZER
Product CodeDKH
Date Received2012-03-29
Catalog Number6801375
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerORTHO-CLINICAL DIAGNOSTICS
Manufacturer Address100 INDIGO CREEK DRIVE ROCHESTER NY 14626 US 14626


Patients

Patient NumberTreatmentOutcomeDate
10 2012-03-29

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.