VITROS 250 CHEMISTRY SYSTEM 8132086

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2012-04-03 for VITROS 250 CHEMISTRY SYSTEM 8132086 manufactured by Ortho-clinical Diagnostics.

Event Text Entries

[14929718] A customer observed a lower than expected vitros phyt result (<3. 0 ug/ml) on a single proficiency fluid while using the vitros 250 chemistry system. The proficiency fluid's expected result was 6. 69 ug/ml. Biased results of the magnitude and direction observed may lead to inappropriate physician action if patient samples were affected. The customer made no allegations that patient sample results were affected. There was no report of patient harm as a result of this event. This report corresponds to ortho clinical diagnostics inc. (b)(4).
Patient Sequence No: 1, Text Type: D, B5


[15294916] The investigation found no evidence to suggest that either the vitros 250 or the vitros phyt slides malfunctioned. The investigation determined that a "no result" was generated for phyt on the affected proficiency fluid sample due to the matrix of the fluid. In addition to the "no result", the vitros 250 generated a condition code. The operator did not respond to appropriately to an associated analyzer condition code. When the vitros 250 analyzer displays a generated condition code, an error text is displayed that provides information on how to interpret the analyzer codes. The root cause of the reported lower than expected phyt proficiency fluid result was user error, as the operator incorrectly interpreted the analyzer condition code as indicating a phyt result below the analyzer's reportable range, when in fact a result could not be obtained due to the matrix of the proficiency fluid. The customer was made aware of the improper response to the condition code to avoid additional misinterpretation.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1319681-2012-00053
MDR Report Key2515513
Report Source05
Date Received2012-04-03
Date of Report2012-04-03
Date of Event2012-01-27
Date Mfgr Received2012-03-05
Device Manufacturer Date2000-01-05
Date Added to Maude2012-04-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMR. JOSEPH FALVO
Manufacturer Street100 INDIGO CREEK DRIVE
Manufacturer CityROCHESTER NY 14626
Manufacturer CountryUS
Manufacturer Postal14626
Manufacturer Phone5854533000
Manufacturer G1ORTHO-CLINICAL DIAGNOSTICS
Manufacturer Street100 INDIGO CREEK DRIVE
Manufacturer CityROCHESTER NY 14626
Manufacturer CountryUS
Manufacturer Postal Code14626
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameVITROS 250 CHEMISTRY SYSTEM
Generic NameCHEMISTRY ANALYZER
Product CodeDKH
Date Received2012-04-03
Catalog Number8132086
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerORTHO-CLINICAL DIAGNOSTICS
Manufacturer Address100 INDIGO CREEK DRIVE ROCHESTER NY 14626 US 14626


Patients

Patient NumberTreatmentOutcomeDate
10 2012-04-03

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.