MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 07 report with the FDA on 2012-04-06 for PATCH 0041 manufactured by Historical Puerto Rico.
[20734868]
Boston scientific received information that during a normal device changeout, this epicardial lead patch was noted to be fractured. Fluoroscopy confirmed that there was a break in the patch conductor coils. The lead was surgically abandoned and replaced with no adverse patient effects were reported.
Patient Sequence No: 1, Text Type: D, B5
[20904891]
The lead was surgically abandoned. No further information is available. This report will be updated if more information becomes available.
Patient Sequence No: 1, Text Type: N, H10
Report Number | 2124215-2012-01718 |
MDR Report Key | 2520267 |
Report Source | 07 |
Date Received | 2012-04-06 |
Date of Report | 2012-01-27 |
Date of Event | 2012-01-27 |
Date Mfgr Received | 2012-01-27 |
Date Added to Maude | 2012-04-06 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Reporter Occupation | MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 0 |
Manufacturer Contact | SHARON ZURN |
Manufacturer Street | 4100 HAMLINE AVE. N |
Manufacturer City | ST. PAUL MN 55112 |
Manufacturer Postal | 55112 |
Manufacturer Phone | 6515824786 |
Manufacturer G1 | HISTORICAL PUERTO RICO |
Manufacturer Street | GUIDANT PUERTO RICO B. V. |
Manufacturer City | DORADO PR |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | PATCH |
Generic Name | IMPLANTABLE LEAD |
Product Code | NHW |
Date Received | 2012-04-06 |
Model Number | 0041 |
ID Number | L 67 |
Device Expiration Date | 1997-06-21 |
Operator | LAY USER/PATIENT |
Device Availability | N |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | HISTORICAL PUERTO RICO |
Manufacturer Address | GUIDANT PUERTO RICO B. V. DORADO PR |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2012-04-06 |