[20000294] It was reported that the sleeve section of the pt helper bent where it attaches to the bed bracket. Additional clinical follow up with the hospital indicated that the pt helper had been on the pt's bed providing trapeze use for quite some time. The rn stated the pt was considered to be quite proactive with physical therapy and self exercise to regain limb function and strength. The rn stated the pt was "doing chin-ups on the trapeze, lifting his full body weight completely off the bed, when the pt noted the bar to be bending slightly over the bed. " the rn stated the pt was not harmed and neither the trapeze no the pt helper fell.
Patient Sequence No: 1, Text Type: D, B5


[20279322] The device was not returned to the manufacturer at the time of this report. A follow up medwatch will be submitted once the device has been returned and the investigation is complete.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1035617-2012-00002
MDR Report Key2521021
Report Source05,08
Date Received2012-04-04
Date of Report2012-03-07
Date of Event2012-03-01
Date Mfgr Received2012-03-07
Date Added to Maude2012-09-07
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactCINDY DICKEY
Manufacturer Street1872 OLD MOUNTAIN RD.
Manufacturer CitySTATESVILLE NC 28625
Manufacturer CountryUS
Manufacturer Postal28625
Manufacturer Phone7048731001
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePATIENT HELPER ADAPTER (HEIGHT ADJ)
Generic NamePATIENT HELPER ADAPTER (HEIGHT ADJ)
Product CodeILZ
Date Received2012-04-04
Model NumberNA
Catalog Number00-0965-002-01
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerZIMMER SURGICAL
Manufacturer Address1872 OLD MOUNTAIN RD. STATESVILLE NC 28625 US 28625


Patients

Patient NumberTreatmentOutcomeDate
10 2012-04-04

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.