"FLOWMETER" OR MEDICAL GAS CONNECTOR

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 99 report with the FDA on 2010-06-29 for "FLOWMETER" OR MEDICAL GAS CONNECTOR manufactured by Chemtron (distributed By Allied Health Care).

Event Text Entries

[2584277] Oxygen or air fittings disconnect at knurled fitting causing flowmeter to project from wall. This has the potential to injure users or pts. It also has the potential to discontinue the flow of oxygen to a pt who needs oxygen to survive.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2527248
MDR Report Key2527248
Report Source99
Date Received2010-06-29
Date of Report2010-06-07
Date Facility Aware2010-06-04
Report Date2010-06-07
Date Reported to FDA2010-06-17
Date Added to Maude2012-04-13
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use0
Previous Use Code3
Event Type3
Type of Report3

Device Sequence Number: 1

Brand Name"FLOWMETER" OR MEDICAL GAS CONNECTOR
Generic Name"FLOWMETER" OR MEDICAL GAS CONNECTOR
Product CodeCAX
Date Received2010-06-29
Device AvailabilityY
Device Age4 YR
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerCHEMTRON (DISTRIBUTED BY ALLIED HEALTH CARE)

Device Sequence Number: 2

Brand Name"FLOWMETER" OR MEDICAL GAS CONNECTOR
Generic Name"FLOWMETER" OR MEDICAL GAS CONNECTOR
Product CodeCAX
Date Received2010-06-29
Device AvailabilityY
Device Eval'ed by Mfgr*
Device Sequence No2
Device Event Key0
ManufacturerSUPERIOR PRODUCTS (DISTRIBUTED BY ALLIED HEALTH CARE)


Patients

Patient NumberTreatmentOutcomeDate
101. Life Threatening 2010-06-29

© 2026 FDA.report