CLEAR-PB MOBILE BARRIERS 56-604

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 07 report with the FDA on 2012-04-09 for CLEAR-PB MOBILE BARRIERS 56-604 manufactured by Fluke Biomedical.

Event Text Entries

[2608038] Customer claims 11 yr old (b)(4) clear pb mobile barrier broke during use. Complaint description as provided in customer supplied 3500a: "event desc: the pt was undergoing coronary angiography. The attending physician was walking across the procedure room in order to insert some documents into the pt's chart. Since fluoroscopy was on going, he was moving the mobile x-ray shield across the room so that he would remain behind it. Just after he let go, the upper portion of the window snapped off just above the vertical mounts on both sides. The cracked edge was smooth but not sharp like glass. The piece that fell to the floor did not shatter nor strike anyone or anything that would have compromised the procedure. Staff fortunately saved the two parts of the broken window. It's made of plastic, not glass so the part that fell didn't shatter. The device is considered a radiation safety device by the mfr. " "my opinion is that there was a notch near one of the side support channels and that the continued flexing and handling of the unsupported upper portion of the window over a long period of time resulted in a crack growing from the notch until this catastrophic event occurred. "
Patient Sequence No: 1, Text Type: D, B5


[9954364] Product was mfg by nuclear associates who was subsequently acquired by fluke biomedical. Customer has been contacted to determine if product is available for investigation. No response received as of apr 5, 2012. F/u report will be submitted when additional info is received.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2921581-2012-00001
MDR Report Key2538070
Report Source07
Date Received2012-04-09
Date of Report2012-03-26
Date of Event2012-02-27
Date Mfgr Received2012-03-26
Date Added to Maude2012-09-11
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional0
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactJOHN NELSON
Manufacturer Street6920 SEAWAY BLVD.
Manufacturer CityEVERETT WA 98203
Manufacturer CountryUS
Manufacturer Postal98203
Manufacturer Phone4404982579
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCLEAR-PB MOBILE BARRIERS
Generic NameSHIELD, X-RAY MOBILE
Product CodeKPY
Date Received2012-04-09
Model Number56-604
Catalog Number56-604
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerFLUKE BIOMEDICAL
Manufacturer Address6920 SEAWAY BLVD. EVERETT WA US


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2012-04-09

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.