COULTER? LH 750 SLIDEMAKER 6605633

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,05,06 report with the FDA on 2012-04-17 for COULTER? LH 750 SLIDEMAKER 6605633 manufactured by Beckman Coulter Inc..

Event Text Entries

[2607546] A customer contacted beckman coulter inc. (bec) reporting that their coulter lh 750 slidemaker had a blood leak that was contained within the instrument. The operators were wearing personal protective equipment (ppe) consisting of gloves, gown and goggles at the time of the incident. No injury or exposure was reported. There was no impact to patient samples.
Patient Sequence No: 1, Text Type: D, B5


[9956203] A field service engineer (fse) went on-site the day of the event and stated that no leak was present at the time of site visit, but cleaned up signs of old leak. Per fse, a technician had slides piled up under the dispense probe area that were pushed by the moving truck due to which rinse block (rb1) came out of the mounting brackets. The technician reinstalled rinse block rb1 which resolved the issue but did not clean up the spilled fluids below the rinse block. Another technician later discovered the fluids and initiated service call. The cause of the leak was the rinse block (rb1) that was pushed out of mounting brackets by a pile up of slides. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1061932-2012-01189
MDR Report Key2538631
Report Source01,05,06
Date Received2012-04-17
Date of Report2012-03-22
Date of Event2012-03-22
Date Mfgr Received2012-03-22
Device Manufacturer Date2010-06-01
Date Added to Maude2012-09-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS. DUNG NGUYEN
Manufacturer Street250 S KRAEMER BLVD.
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal92821
Manufacturer Phone7149614941
Manufacturer G1BECKMAN COULTER INC.
Manufacturer Street11800 SW 147TH AVENUE,
Manufacturer CityMIAMI FL 33196
Manufacturer CountryUS
Manufacturer Postal Code33196
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameCOULTER? LH 750 SLIDEMAKER
Generic NameAUTOMATED SLIDE SPINNER
Product CodeGKJ
Date Received2012-04-17
Model NumberLH 750 SLIDEMAKER
Catalog Number6605633
Lot NumberNA
ID NumberSW VERSION 2D1
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBECKMAN COULTER INC.
Manufacturer Address250 S KRAEMER BLVD. BREA CA 92821 US 92821


Patients

Patient NumberTreatmentOutcomeDate
10 2012-04-17

© 2026 FDA.report