UNICEL? DXH SLIDEMAKER STAINER N/A A82669

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,05,06 report with the FDA on 2012-04-20 for UNICEL? DXH SLIDEMAKER STAINER N/A A82669 manufactured by Beckman Coulter Inc..

Event Text Entries

[22111078] The customer reported that bath 1 of the unicel dxh slidemaker stainer had overflowed with methanol into the bath tray. The customer indicated that the bath tray filled completely and ultimately overflowed into the instrument. Approximately 10 ml was spilled into the instrument. The customer was wearing personal protective equipment consisting of gloves, lab coat and eye protection and no exposure or injury was reported. Patient results were not affected and there was no change to patient treatment in connection with this event. Beckman coulter field service engineer (fse) stated that the bath 1 sensor and overflow sensor were covered with debris causing the dxh slidemaker stainer to overflow. Fse proceeded to replace the bath 1 and overflow sensors and repair was verified per established procedures.
Patient Sequence No: 1, Text Type: D, B5


[22132030] (b)(4).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1061932-2012-01262
MDR Report Key2544072
Report Source01,05,06
Date Received2012-04-20
Date of Report2012-03-28
Date of Event2012-03-28
Date Mfgr Received2012-03-28
Device Manufacturer Date2012-01-02
Date Added to Maude2012-09-13
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationMEDICAL TECHNOLOGIST
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS. DUNG NGUYEN
Manufacturer Street250 S. KRAEMER BOULEVARD
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal92821
Manufacturer Phone7149614941
Manufacturer G1BECKMAN COULTER, INC.
Manufacturer Street11800 SW 147TH AVENUE
Manufacturer CityMIAMI FL 33196
Manufacturer CountryUS
Manufacturer Postal Code33196
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameUNICEL? DXH SLIDEMAKER STAINER
Generic NameSPINNER, SLIDE, AUTOMATED
Product CodeGKJ
Date Received2012-04-20
Model NumberN/A
Catalog NumberA82669
ID NumberSW VERSION 1.5.1.0
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBECKMAN COULTER INC.
Manufacturer Address250 S. KRAEMER BLVD BREA CA 92821 US 92821


Patients

Patient NumberTreatmentOutcomeDate
10 2012-04-20

© 2026 FDA.report