TRITON SPINE THERAPY TABLE TRT600

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 08 report with the FDA on 2012-04-06 for TRITON SPINE THERAPY TABLE TRT600 manufactured by Djo, Llc.

Event Text Entries

[2809619] No pts were involved, the sales rep was performing an in service at the customer's office, he pulled down the hand lever, the weld broke and the smaller chest/head section broke. Initial info provided to djo on (b)(6) 2012, was insufficient in that it did not suggest that device failed to meet specifications or malfunctioned. The initial info provided did not suggest that if the alleged malfunction or failure to meet specifications were to recur it would cause or contribute to death or serious injury. Photos of the broken weld were sent to djo. The photos revealed that the weld under the chest/head section (which supports the pt's upper body including neck and back) is broken. Although a pt was not involved, the photos reveal that a malfunction of the device occurred, and suggest that would the malfunction were to recur, it can possibly cause or contribute to a death or serious injury. Given the new info djo has classified the product problem as a reportable.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9616086-2012-00021
MDR Report Key2545968
Report Source08
Date Received2012-04-06
Date of Report2012-04-05
Date Mfgr Received2012-03-07
Date Added to Maude2012-09-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer Street1430 DECISION ST
Manufacturer CityVISTA CA 92081
Manufacturer CountryUS
Manufacturer Postal92081
Manufacturer Phone7607271280
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTRITON SPINE THERAPY TABLE
Generic NameTRACTION POWERED -TABLES AND TRACTION
Product CodeIRS
Date Received2012-04-06
Model NumberTRT600
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerDJO, LLC
Manufacturer AddressVISTA CA US


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2012-04-06