SS-2 120V W/O PKS 2304

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2008-09-29 for SS-2 120V W/O PKS 2304 manufactured by Chattanooga Group.

Event Text Entries

[2747691] A ss-2 hydrocollator hot pack heating unit overheated. The unit began to smoke and emit flames from under the heating tank lid. An employee of the facility contacted the fire dept at 6:24 am. The fire dept was dispatched, as a non-emergency dispatch, to the facility. The fire dept removed the device from the facility. The device tank was filled with water to extinguish the flames. The fire dept noted a small amount of scorching on the facility wall where the device had been positioned. The facility had a light haze of smoke, thus ventilation fans were positioned to remove the smoke. The ventilation was operated until 7:15 am. The hydrocollator was purchased in (b)(6) 2008. Upon placing in service the device would malfunction by boiling the tank water. A call was made to the distributor requesting a replacement and the distributor suggested replacing the thermostat. A replacement thermostat was sent to the customer and the physical therapist replaced the thermostat. The client said that the unit continued to boil and they were concerned. Another call was made requesting a replacement hydrocollator just before the fire occurred. The water level in the hydrocollator was checked the day before.
Patient Sequence No: 1, Text Type: D, B5


[9965736] The device has been received and is currently under eval. The device eval and any add'l findings will be provided upon conclusion of the device eval.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1022819-2008-00265
MDR Report Key2555645
Report Source05
Date Received2008-09-29
Date of Report2008-09-04
Date of Event2008-09-04
Date Mfgr Received2008-09-05
Date Added to Maude2012-05-03
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMICHAEL TREAS
Manufacturer Street4717 ADAMS RD.
Manufacturer CityHIXSON TN 37343
Manufacturer CountryUS
Manufacturer Postal37343
Manufacturer Phone4238702281
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSS-2 120V W/O PKS
Product CodeIRQ
Date Received2008-09-29
Returned To Mfg2008-09-15
Model Number2304
Catalog Number2304
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerCHATTANOOGA GROUP
Manufacturer AddressHIXSON TN US


Patients

Patient NumberTreatmentOutcomeDate
10 2008-09-29