OLYMPUS VIDEO TELESCOPE A50021A

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2012-04-24 for OLYMPUS VIDEO TELESCOPE A50021A manufactured by Olympus Winter & Ibe Gmbh.

Event Text Entries

[2750550] The user facility reported that during a therapeutic laparoscopic cholecystectomy procedure, the users experienced a complete loss of image. The procedure was completed using a different device. There was no pt injury report.
Patient Sequence No: 1, Text Type: D, B5


[9833914] The device referenced in this report was returned to olympus for eval. The eval was able to duplicate the user's report of image difficulty. The image flickered, distorted and then the image was lost when the video cable was manipulated. The eval revealed that the outer tube (r-unit) of the telescope was bent, dented at the distal end and there were minor scratches on the objective lens. The cable had a damaged signal wire at the r-unit connection section. The cause of the image difficulty was likely due to physical damage to the cable unit and outer tube, which is related user handling. The device was serviced and returned to the user facility. This report is being submitted as a medical device report in an excess of caution.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9610773-2012-00019
MDR Report Key2556642
Report Source06
Date Received2012-04-24
Date of Report2012-03-26
Date of Event2012-03-01
Date Mfgr Received2012-03-26
Date Added to Maude2012-09-13
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactLAURA STORMS-TYLER
Manufacturer Street2400 RINGWOOD AVE.
Manufacturer CitySAN JOSE CA 95131
Manufacturer CountryUS
Manufacturer Postal95131
Manufacturer Phone4848965688
Manufacturer G1OLYMPUS WINTER & IBE GMBH
Manufacturer StreetKUEHNSTRASSE 61
Manufacturer CityHAMBURG 22045
Manufacturer CountryGM
Manufacturer Postal Code22045
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameOLYMPUS VIDEO TELESCOPE
Generic NameVIDEO TELESCOPE
Product CodeFBP
Date Received2012-04-24
Returned To Mfg2012-03-26
Model NumberA50021A
Catalog NumberA50021A
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerOLYMPUS WINTER & IBE GMBH
Manufacturer AddressKUEHNSTRASSE 61 HAMBURG 22045 GM 22045


Patients

Patient NumberTreatmentOutcomeDate
10 2012-04-24

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.