CELLEBRITY? M00516071 1607

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 00,01,05,07 report with the FDA on 2012-05-03 for CELLEBRITY? M00516071 1607 manufactured by Boston Scientific - Spencer.

Event Text Entries

[2643133] It was reported to boston scientific corporation that a celebrity cytology brush was used during a pulmonary biopsy procedure. According to the complainant, when the user attempted to retract the device after brushing the bronchi, the pull wire tore near the handle inside the scope. No pieces detached inside the patient. Since the handle and pull wire were separated, it was reported that the pull wire probably remained inside the scope after being torn. The procedure was completed with this device. There were no patient complications reported as a result of this event. The patient's condition at the conclusion of the procedure was reported to be good.
Patient Sequence No: 1, Text Type: D, B5


[9890165] Visual evaluation of the returned device found that the pull wire had separated from the handle cannula; the pull wire had been completely removed from the sheath. The pull wire was bent near the distal tip, and the outer sheath was kinked in several locations. The handle cannula functioned smoothly. The handle was then disassembled and it was noted that there were crimp marks present on all four sides of the handle cannula and scratches on the proximal end of the pull wire, indicating proper crimping during manufacturing. Additionally, the proximal end of the pull wire was fed back into the handle cannula, but it would not pass through the crimped area. The complaint was confirmed; the pull wire had separated from the handle cannula, resulting in a "torn" appearance. As evidence was found that the device was assembled properly during manufacturing, it is more likely that excessive force was applied to the device, subsequently causing the separation. Therefore, the most probable root cause of the defects identified is operational context. A review of the device history record (dhr) was performed; no anomalies were noted. A search of the complaint database revealed that no other complaints exist for the specified lot.
Patient Sequence No: 1, Text Type: N, H10


[21356428] Event date is unknown. Although the suspect device has been received, the evaluation has not been completed. Therefore, the cause of the reported malfunction has not been determined. Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
Patient Sequence No: 1, Text Type: N, H10


[21412872] It was reported to boston scientific corporation that a celebrity cytology brush was used during a pulmonary biopsy procedure. According to the complainant, when the user attempted to retract the device after brushing the bronchi, the pull wire tore near the handle inside the scope. No pieces detached inside the patient. Since the handle and pull wire were separated, it was reported that the pull wire probably remained inside the scope after being torn. The procedure was completed with this device. There were no patient complications reported as a result of this event. The patient's condition at the conclusion of the procedure was reported to be good.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3005099803-2012-01627
MDR Report Key2558639
Report Source00,01,05,07
Date Received2012-05-03
Date of Report2012-04-10
Date Mfgr Received2012-05-10
Device Manufacturer Date2011-07-13
Date Added to Maude2012-05-03
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactEMP. CHARLES MONTGOMERY
Manufacturer Street100 BOSTON SCIENTIFIC WAY
Manufacturer CityMARLBOROUGH MA 01752
Manufacturer CountryUS
Manufacturer Postal01752
Manufacturer Phone5086834000
Manufacturer G1BOSTON SCIENTIFIC - SPENCER
Manufacturer Street780 BROOKSIDE DRIVE
Manufacturer CitySPENCER IN 47460
Manufacturer CountryUS
Manufacturer Postal Code47460
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCELLEBRITY?
Generic NameBRUSH, BIOPSY, BRONCHOSCOPE (NON-RIGID)
Product CodeBTG
Date Received2012-05-03
Returned To Mfg2012-04-19
Model NumberM00516071
Catalog Number1607
Lot Number14500539
Device Expiration Date2013-07-10
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBOSTON SCIENTIFIC - SPENCER
Manufacturer Address780 BROOKSIDE DRIVE SPENCER IN 47460 US 47460


Patients

Patient NumberTreatmentOutcomeDate
10 2012-05-03

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.