IMMAGE? 800 IMMUNOCHEMISTRY SYSTEM IMMAGE 800 A15445

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2012-05-16 for IMMAGE? 800 IMMUNOCHEMISTRY SYSTEM IMMAGE 800 A15445 manufactured by Beckman Coulter, Inc..

Event Text Entries

[21765024] Customer stated that the wash station overflowed when attempted to wash the probe on the immage 800 immunochemistry systems. Leaking hardware component could potentially impact sample results by either contamination or dilution effect. The customer was wearing personal protective equipment (ppe) consisting of a lab coat and gloves. The customer did not come in contact with the fluid and did not seek medical attention. No exposure (sprayed or splashed) to mucous membranes or open wounds were reported. No death, injury or changes to patient treatment or results are associated with this event. The customer reviewed the msds and the facility has an exposure control or risk management plan in place. On (b)(6) 2012, a field service engineer ((fse) found sample probe wash station overflowed due to clogged waste filter. The fse replaced the filter which resolved issue. The cause of the leak was due to clogged waste filter.
Patient Sequence No: 1, Text Type: D, B5


[21924542] (b)(4).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2050012-2012-01072
MDR Report Key2575541
Report Source05,06
Date Received2012-05-16
Date of Report2012-04-17
Date of Event2012-04-17
Date Mfgr Received2012-04-17
Device Manufacturer Date2008-09-05
Date Added to Maude2012-09-24
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS. DUNG NGUYEN
Manufacturer Street250 S. KRAEMER BLVD.
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal92821
Manufacturer Phone7149614941
Manufacturer G1BECKMAN COULTER, INC.
Manufacturer Street250 S KRAEMER BLVD
Manufacturer CityBREA, CA 92821
Manufacturer CountryUS
Manufacturer Postal Code92821
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameIMMAGE? 800 IMMUNOCHEMISTRY SYSTEM
Generic NameNEPHELOMETER, FOR CLINICAL USE
Product CodeCFN
Date Received2012-05-16
Model NumberIMMAGE 800
Catalog NumberA15445
Lot NumberN/A
ID NumberN/A
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBECKMAN COULTER, INC.
Manufacturer Address250 S KRAEMER BLVD BREA, CA 92821 US 92821


Patients

Patient NumberTreatmentOutcomeDate
10 2012-05-16

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.