CHOLESTECH LDX NORTH AMERICA 14-874

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2012-05-02 for CHOLESTECH LDX NORTH AMERICA 14-874 manufactured by Alere San Diego, Inc..

Event Text Entries

[2649747] Customer reported that ldx analyzer would constantly overheat and be hot to touch. Customer also added that tray door opens when it shouldn't and often gets a mag. Read error/used cassette message. She did not say they were burned or injured in any way by the overheating. Technical services replaced customer's meter.
Patient Sequence No: 1, Text Type: D, B5


[9861741] Product support investigated returned ldx analyzer for customer's claims of overheating. Six lipid cassettes were run on the returned ldx analyzer and no overheating occurred. Six alt/ast cassettes were also run on the analyzer and no overheating occurred. No temperature out of range messages were obtained and the analyzer never felt warm or hot to the touch during testing. The analyzer's temperature readings were between 26c and 28c during the alt/ast testing. No errors were obtained during testing and the analyzer functioned as expected with no problems found. The customer's complaints of overheating, tray opening when it shouldn't, mag. Read errors and used cassette errors were not replicated. No corrective action required at this time as no product deficiency was established.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2027969-2012-00633
MDR Report Key2575642
Report Source05
Date Received2012-05-02
Date of Report2012-05-02
Date of Event2012-04-05
Date Mfgr Received2012-04-05
Date Added to Maude2012-06-13
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactCARMEN BERGELIN, MANAGER
Manufacturer Street9975 SUMMERS RIDGE RD
Manufacturer CitySAN DIEGO CA 92121
Manufacturer CountryUS
Manufacturer Postal92121
Manufacturer Phone8588052256
Single Use3
Previous Use Code3
Removal Correction NumberNI
Event Type3
Type of Report3

Device Details

Brand NameCHOLESTECH LDX NORTH AMERICA
Generic NameCHOLESTEROL METER
Product CodeCHH
Date Received2012-05-02
Returned To Mfg2012-04-16
Model Number14-874
Lot NumberNG
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerALERE SAN DIEGO, INC.
Manufacturer AddressSAN DIEGO CA US


Patients

Patient NumberTreatmentOutcomeDate
10 2012-05-02

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.