BACT/ALERT BPA AND BPN CULTURE BOTTLES 279018 279019

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2008-12-09 for BACT/ALERT BPA AND BPN CULTURE BOTTLES 279018 279019 manufactured by Biomerieux, Inc..

Event Text Entries

[2719866] A recipient at the (b)(6) in (b)(6), received a partially transfused platelets pheresis unit and experienced severe dyspnea and an elevated temperatures of 40 degrees c.
Patient Sequence No: 1, Text Type: D, B5


[9860895] An investigation of this incident has been initiated by biomerieux inc. The community blood center initiated a transfusion reaction investigation on (b)(6) 2008. The report stated that "the implicated unit was collected on (b)(6) 2008 and sterility sampling was performed on (b)(6) 2008. The bact/alert bottles remained negative and were removed from the incubator on (b)(6) 2008. The aerobic and anaerobic bottles along with the associated fresh frozen plasma unit were sent to the reference lab for cultures; no bacterial growth was detected. " "the gram stain and culture from the partially transfused platelet product showed gram positive cocci and the culture grew (b)(6). " "investigation reveals the bact/alert system for notification was functioning and review by the fda indicated that all standard operating procedure had been followed. "
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3002769706-2008-00009
MDR Report Key2576515
Report Source05,06
Date Received2008-12-09
Date of Report2008-12-03
Date of Event2008-10-14
Date Mfgr Received2008-11-11
Device Manufacturer Date2008-05-01
Date Added to Maude2012-05-21
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactELIZABETH FRITZ
Manufacturer Street100 RODOLPHE ST
Manufacturer CityDURHAM NC 27712
Manufacturer CountryUS
Manufacturer Postal27712
Manufacturer Phone9196202682
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameBACT/ALERT BPA AND BPN CULTURE BOTTLES
Generic NameMICROBIAL GROWTH MONITOR
Product CodeJTA
Date Received2008-12-09
Catalog Number279018 279019
Lot Number1020387 1020417
Device Expiration Date2009-05-30
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBIOMERIEUX, INC.
Manufacturer Address100 RODOLPHE ST DURHAM NC US


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2008-12-09

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.