COBAS 6000 C501 MODULE 05860636001

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2012-05-18 for COBAS 6000 C501 MODULE 05860636001 manufactured by Roche Diagnostics.

Event Text Entries

[2709845] The customer had been receiving questionable gentamicin results on their c501 analyzer for the past few weeks. The customer provided data for six samples, of which one had discrepant results reported outside the laboratory. All testing was performed on the same c501 analyzer. The patient's initial gentamicin result was 0 ug/ml. On (b)(6) 2012, the repeat result was 1. 02 ug/ml. The customer considered the repeat result to be correct. It was unknown if the patients were adversely affected. The gentamicin reagent lot number was 64602901 and the expiration date was 10/31/2012. The field service representative determined the sample probe had gel on the bottom of the probe. He replaced the sample probe. Calibrations and quality controls were run with results within range.
Patient Sequence No: 1, Text Type: D, B5


[9885011] It was unknown if the initial reporter sent report to the fda.
Patient Sequence No: 1, Text Type: N, H10


[18750873] .
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1823260-2012-02648
MDR Report Key2580329
Report Source05,06
Date Received2012-05-18
Date of Report2012-08-10
Date of Event2012-04-28
Date Mfgr Received2012-05-07
Date Added to Maude2012-05-18
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location3
Manufacturer ContactNA JENNIFER WOLFGRAM
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175217008
Manufacturer G1HITACHI HIGH TECH CORP.
Manufacturer Street882 ICHIGE HITACHINAKA NA
Manufacturer CityIBARAKI 312-8504
Manufacturer Postal Code312-8504
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameCOBAS 6000 C501 MODULE
Generic NameCLINICAL CHEMISTRY ANALYZER
Product CodeLCD
Date Received2012-05-18
Model NumberNA
Catalog Number05860636001
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 46250 US 46250