TEN20 CONDUCTIVE PASTE 10-20-XX

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2012-05-16 for TEN20 CONDUCTIVE PASTE 10-20-XX manufactured by D.o. Weaver And Company.

Event Text Entries

[2720946] Patient had two eegs conducted, the first in (b)(6) 2011, and the second in (b)(6) 2011. The pt reacted negatively in both instances, but more severly in the first test. The first test, ten20 conductive paste, nuprep skin prep gel, and gold electrodes were used. The electrodes were glued in place with collodion. This test lasted 3 days. After the test, the pt had a severe skin reaction/redness at the electrode sites. During the second test, because the pt reacted negatively when nuprep and ten20 were used, the lab used different products for the test. The pt reacted to this test with redness as well, but not as severe as the first test.
Patient Sequence No: 1, Text Type: D, B5


[9891732] Lab tech believes pt likely has highly sensitive skin and possibly a "metal allergy" (since nuprep contains aluminum oxide). Reporter also stated that the technologist who administered the test is known to be fairly aggressive when prepping the skin. Because nuprep is abrasive, this may have contributed to the severe reaction in the first test. It is uncertain which products (nuprep, ten20, or the electrodes) caused the reaction.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1718791-2012-00009
MDR Report Key2580959
Report Source05
Date Received2012-05-16
Date of Report2012-05-16
Date of Event2011-10-01
Date Added to Maude2012-05-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactNICHOLAS LEE
Manufacturer Street565 NUCLA WAY, UNIT B
Manufacturer CityAURORA CO 80011
Manufacturer CountryUS
Manufacturer Postal80011
Manufacturer Phone3033661804
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTEN20 CONDUCTIVE PASTE
Generic NameELECTROCONDUCTIVE MEDIA (CODE: GYB)
Product CodeGYB
Date Received2012-05-16
Model Number10-20-XX
Catalog Number10-20-XX
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerD.O. WEAVER AND COMPANY
Manufacturer AddressAURORA CO US


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2012-05-16

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.