OLYMPUS EVIS EXERA COLONOVIDEOSCOPE CF-Q160AL

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2012-05-18 for OLYMPUS EVIS EXERA COLONOVIDEOSCOPE CF-Q160AL manufactured by Olympus Medical Systems Corporation.

Event Text Entries

[2659230] The user facility reported that during a diagnostic colonoscopy the users experienced a complete loss of image. The colonoscope was withdrawn from the pt and the procedure was completed using a different but similar endoscope. There was no pt injury reported.
Patient Sequence No: 1, Text Type: D, B5


[9888070] The device referenced in this report was returned to olympus for eval. The eval did not confirm the user's report of image loss; however, there was evidence of fluid invasion in the electrical connector, scope connector, and control body unit, which likely caused the image difficulty. The device passed the leak test. The bending section cover and bending section cover glue are found to be a non-olympus repairs. The objective lens was cracked. The switch #1 button is not working. The insertion tube failed the insulation test. The device was refurbished. This report is being submitted as a medical device report in an excess of caution.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number8010047-2012-00159
MDR Report Key2588945
Report Source06
Date Received2012-05-18
Date of Report2012-04-20
Date of Event2012-04-01
Date Mfgr Received2012-04-20
Date Added to Maude2012-06-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactLAURA STORMS-TYLER
Manufacturer Street2400 RINGWOOD AVE
Manufacturer CitySAN JOSE CA 95131
Manufacturer CountryUS
Manufacturer Postal95131
Manufacturer Phone4848965688
Manufacturer G1OLYMPUS MEDICAL SYSTEMS CORPORATION
Manufacturer Street2951 ISHIKAWA-CHO
Manufacturer CityHACHIOJI-SHI, TOKYO 192-8507
Manufacturer CountryJA
Manufacturer Postal Code192-8507
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameOLYMPUS EVIS EXERA COLONOVIDEOSCOPE
Generic NameCOLONOSCOPE
Product CodeFTJ
Date Received2012-05-18
Returned To Mfg2012-04-20
Model NumberCF-Q160AL
Catalog NumberCF-Q160AL
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerOLYMPUS MEDICAL SYSTEMS CORPORATION
Manufacturer Address2951 ISHIKAWA-CHO HACHIOJI-SHI, TOKYO 192-8507 JA 192-8507


Patients

Patient NumberTreatmentOutcomeDate
10 2012-05-18

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.