CENTURY SAF-KARY SK-10 N/A

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 1995-01-19 for CENTURY SAF-KARY SK-10 N/A manufactured by Century Mfg. Co..

Event Text Entries

[14401] Nursing assistant was transporting resident to tub via tub chair with safety belt applied. Frame of chair broke and seat of chair fell to floor. Resident struck left knee on floor and right shoulder on wall. Sustained ecchymotic area and superficial abrasion to left knee. Area cleansed and ice applied. Right shoulder and left knee were x-rayed on 5/5/94 and both x-rays negative. Upon inspection of chair rust was identified on frame where frame broke. Device not labeled for single use. Patient medical status prior to event: satisfactory condition. There was not multiple patient involvement. Device serviced in accordance with service schedule. Date last serviced: 01-apr-94. Service provided by: other. Service records not available. No imminent hazard to public health claimed. Device used as labeled/intended. Device was evaluated after the event. Method of evaluation: actual device involved in incident was evaluated, visual examination. Results of evaluation: component failure, material degradation/deterioration, component failure. Conclusion: device failure occurred and was related to event, device failure directly caused event. Certainty of device as cause of or contributor to event: yes. Corrective actions: device discarded. The device was destroyed/disposed of.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number25898
MDR Report Key25898
Date Received1995-01-19
Date of Report1995-01-02
Date of Event1994-05-05
Date Facility Aware1994-05-05
Report Date1995-01-02
Date Reported to Mfgr1994-05-13
Date Added to Maude1995-09-20
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional0
Initial Report to FDA0
Report to FDA3
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameCENTURY SAF-KARY
Generic NameN/A
Product CodeIQB
Date Received1995-01-19
Model NumberSK-10
Catalog NumberN/A
ID NumberPAT 3270353
OperatorOTHER HEALTH CARE PROFESSIONAL
Device AvailabilityN
Implant FlagN
Device Sequence No1
Device Event Key26556
ManufacturerCENTURY MFG. CO.


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 1995-01-19