Event Text Entries

[2791164] Female pt underwent nerve conduction study at unreported clinic. Pt reported pulsing sensations after the study was complete.
Patient Sequence No: 1, Text Type: D, B5


[9929910] Pt convinced exam did neurological damage to her system. Pt complains to have tingling sensation in her fingers at times, "pulsing sensation" in her thumb, and tingling sensations in her face. Doctor substituted ten20 paste for the ultrasound gel that is normally used for the procedure. Ten20 is labeled as "not recommended for use in current-inducing electrodes".
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1718791-2012-00011
MDR Report Key2593032
Report Source04
Date Received2012-05-24
Date of Report2012-05-23
Date of Event2012-04-03
Date Mfgr Received2012-05-09
Date Added to Maude2012-06-04
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactNICHOLAS LEE
Manufacturer Street565 NUCLA WAY, UNIT B
Manufacturer CityAURORA CO 80011
Manufacturer CountryUS
Manufacturer Postal80011
Manufacturer Phone3033661804
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTEN20 CONDUCTIVE PASTE
Generic NameELECTROCONDUCTIVE MEDIA (CODE: GYB)`
Product CodeGYB
Date Received2012-05-24
Model Number10-20-XX
Catalog Number10-20-XX
Lot NumberUNK
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerD.O. WEAVER AND COMPANY
Manufacturer AddressAURORA CO US


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2012-05-24

© 2026 FDA.report