VEST MODEL 105 NA

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2012-05-25 for VEST MODEL 105 NA manufactured by Hill-rom Services Private Limited.

Event Text Entries

[2643790] Hill-rom has received a report alleging that a pt believed the best may have contributed to an injury to her aorta where it split into three layers; however, the pt did not know if her belief was correct or not and her doctor did not tell her that the vest had contributed to the injury. Subsequent investigation revealed that her aorta was originally operated on years prior in (b)(6) 2008, and follow up with the clinical manager who spoke with the pt's doctor revealed that the doctor felt it was possible that the best may have been a contributing factor. The information from the clinical manager reasonably suggests that the vest may have contributed to the aorta injury. No device malfunction was alleged nor discovered through subsequent analysis.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3008145987-2012-00001
MDR Report Key2595793
Report Source05
Date Received2012-05-25
Date of Report2012-05-03
Date of Event2012-01-07
Date Mfgr Received2012-05-03
Device Manufacturer Date2011-11-23
Date Added to Maude2012-06-04
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactBRICE RICKER
Manufacturer Street4349 CORPORATE DRIVE
Manufacturer CityCHARLESTON SC 29405
Manufacturer CountryUS
Manufacturer Postal29405
Manufacturer Phone8437408440
Manufacturer G1HILL-ROM SERVICES PRIVATE LIMITED
Manufacturer Street1 YISHUN AVENUE 7
Manufacturer CitySINGAPORE 768923
Manufacturer CountrySN
Manufacturer Postal Code768923
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameVEST MODEL 105
Generic NamePOWERED PERCUSSOR
Product CodeBYI
Date Received2012-05-25
Returned To Mfg2012-04-10
Model Number105
Catalog NumberNA
Lot NumberNA
ID NumberNA
OperatorLAY USER/PATIENT
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerHILL-ROM SERVICES PRIVATE LIMITED
Manufacturer Address1 YISHUN AVE 7 SINGAPORE 768923 SN 768923


Patients

Patient NumberTreatmentOutcomeDate
101. Hospitalization; 2. Life Threatening 2012-05-25

© 2024 FDA.report