[15192346] The customer reported approximately five (5) milliliters of blood and diluent leaked within the instrument involving coulter lh slidemaker. The customer had on proper personal protective equipment (ppe) consisting of gloves, a laboratory coat, and eye protection during the incident. The customer did not have direct contact with the fluid. There was no exposure to open lesions or mucus membranes. There was no report of injury or adverse effect associated with this incident. The customer retested eight (8) patient samples and confirmed patient results were not impacted. There was no patient injury or change in patient treatment associated with this incident. The field service engineer (fse) went to the facility to assess the unit.
Patient Sequence No: 1, Text Type: D, B5


[15752348] The field service engineer (fse) discovered fluid leaked at line vl8 on the aspiration module due to damaged tubing. The fse cleaned the leak area and replaced the tubing on line vl8 in the dispense module. The fse performed fluidic flow checks and noted no further leaks. The fse completed an additional minor preventive maintenance (pm) including slide-making checks, no issues were noted. Service activity performed was verified to meet the specified requirements per established procedures. Results conformed to the manufacturer's published performance specifications. The customer has an exposure control management plan at the facility.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1061932-2012-01804
MDR Report Key2596493
Report Source05,06
Date Received2012-06-01
Date of Report2012-05-15
Date of Event2012-05-15
Date Mfgr Received2012-05-15
Device Manufacturer Date2002-05-01
Date Added to Maude2012-10-03
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS. DUNG NGUYEN
Manufacturer Street250 S. KRAEMER BLVD.
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal92821
Manufacturer Phone7149614941
Manufacturer G1BECKMAN COULTER, INC.
Manufacturer Street11800 SW 147TH AVENUE
Manufacturer CityMIAMI FL 33196
Manufacturer CountryUS
Manufacturer Postal Code33196
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCOULTER? LH SLIDEMAKER
Generic NameSPINNER, SLIDE, AUTOMATED
Product CodeGKJ
Date Received2012-06-01
Model NumberNA
Catalog Number6605633
Lot NumberNA
ID Number2D2
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBECKMAN COULTER, INC.
Manufacturer Address250 S. KRAEMER BLVD. BREA CA 92821 US 92821